临床试验/NCT03323268
NCT03323268已完成不适用

Validation of End-tidal CO2 for Transplumonary Blood Flow Monitoring During PVA-ECMO

University Hospital, Montpellier1 个研究点 分布在 1 个国家目标入组 39 人开始时间: 2016年1月12日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
39
试验地点
1
主要终点
Assessment of end-tidal pressure CO2 (PetCO2) correlation with pulmonary artery flow

研究概览

简要总结

Cardiogenic shock, the most severe form of acute heart failure, is a critical situation where the body cells lack of oxygen because of cardiac dysfunction. The failure of pharmacological therapy to maintain adequate perfusion has led to attempts to improve the circulation by the use of mechanical circulatory support devices. Peripheral veno-arterial support (PVA-ECMO) is the most frequently used device. It consists of extracorporeal circulatory support. Venous blood is aspirated via the right atrium and reinjected into the descending aorta via the femoral artery. As a consequence of this by-pass, pulmonary artery flow and residual left ventricle ejection can fall drastically until zero in the most severe patients. A minimal transpulmonary blood flow is crucial to avoid left heart cavities and pulmonary artery and left heart cavities thrombosis. The gold standard technique to monitor transpulmonary blood flow is right-heart cavities catheterism (Swan-Ganz catheter) but it represents major limits: invasive technique, limited duration of utilization because of septic risk, physical limit of flow measurement (under 1 liter/minute). End-tidal pressure carbon dioxide (Pet CO2) monitoring (or capnography) is a routine and non-invasive measure in ventilated patients. Previous studies have shown that changes in PetCO2 can measure changes in cardiac output in anesthetized patients and that PetCO2 is a useful index of pulmonary artery blood flow during separation from cardiopulmonary bypass. The aim of this study is to demonstrate that PetCO2 is correlated to transpulmonary blood flow in patients under PVA-ECMO and that exhaled CO2 can provide an on-line, continuous, and noninvasive monitor of residual outflow from the heart during PVA-ECMO.

详细描述

Settings

This prospective study is conducted since January 2016 in our tertiary ICU and has been approved by our hospital's institutional review board. Informed consent will be obtained from all patients or their surrogates.

Hemodynamic is monitored via a radial arterial catheter for continuous blood pressure monitoring, a pulmonary artery catheter (Swan-Ganz CCOmbo® CCO/SvO2, Edwards Lifesciences) inserted through the superior vena cava (jugular intern or subclavian vein) and transoesophageal echocardiography (TOE) at PVA-ECMO implantation or at operator discretion.

ECMO circuit settings and patients management under ECMO

PVA-ECMO consist of polyvinyl chloride tubing with a membrane oxygenator (PH.I.S.I.O and EOS; Sorin Group, Clamart, France), a centrifugal pump (Stockert; Sorin Group), and percutaneous or surgically inserted arterial and venous femoral cannulae (Fem-Flex and Fem-Track, Edwards Life- sciences, Guyancourt, France) with or without an additional 7 F cannula inserted distally into the femoral artery to prevent lower limb ischemia. An oxygen-air blender (Sechrist Industries, Anaheim, CA) ventilate the membrane oxygenator. Unfractionated heparin is administrated to maintain an activated partial thromboplastin time between 1.5 and 2 times the normal value. The lower speed flow necessary for adequate tissue perfusion is wanted.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients under mechanical ventilation and PVA-ECMO support with low residual native cardiac output

排除标准

  • age less than 18 years
  • pulmonary disorders with abnormal dead-space (obstructive pulmonary disease; acute respiratory distress syndrome)
  • cardiac shunt (atrial or ventricular communication)
  • significant tricuspid or pulmonary valve disease
  • pulmonary artery catheter implantation contraindicated.

结局指标

主要结局

Assessment of end-tidal pressure CO2 (PetCO2) correlation with pulmonary artery flow

时间窗: Up to 5 days

end-tidal pressure CO2 (PetCO2) vs pulmonary artery flow measured by pulmonary catheter or echocardiography

次要结局

  • Assessment of end-tidal pressure CO2 (PetCO2) thresholds to detect pulmonary artery flow < 1 l/min and indexed flow < 1l/min/m2(Up to 5 days)
  • Assessment of Factors of end-tidal pressure CO2 (PetCO2) variation(Up to 5 days)