ADVANCE-D: ATP Delivery for Painless ICD Therapy

Medtronic Cardiac Rhythm and Heart Failure2 个研究点 分布在 1 个国家目标入组 925 人开始时间: 2004年3月1日最近更新:
适应症

试验速览

阶段
4 期
状态
已完成
发起方
入组人数
925
试验地点
2
主要终点
Efficacy of Anti Tachycardia Pacing (ATP) Therapy to Terminate Fast Ventricular Tachycardia (With Cycle Length of 240ms-320msec)

研究概览

简要总结

The purpose of this study is to estimate and quantify the difference in efficacy of two sequences of ATP therapies (burst 15 pulses, 88% versus burst 8 pulses, 88%) during an episode of spontaneous rhythms classified as fast ventricular tachycardia (FVT) via ventricular fibrillation (VF) in patients who have a Class I or II A indication for ICD implantation, and thus to promote the "painless" therapy aspect of ICD treatment and improve quality of life outcomes for patients.

详细描述

Main objective: Compare & quantify efficacy of two different sequences of burst ATP strategies for the termination of ventricular tachycardia (with Cycle Lenght (CL) of 240ms-320ms) from baseline to 12 months post randomisation in patients who are treated with ATP -8 pulses, 88% versus patients who are treated with ATP -15 pulses, 88%

Secondary objectives:

  • Estimate the efficacy of ATP in successfully treating FVT episodes for patients in primary and secondary prevention
  • Estimate acceleration rate or degeneration of ATP therapy for treating spontaneous FVT episodes in the ATP 15 vs 8 arm
  • Compare the likelihood of syncopal events associated with spontaneous FVT episodes
  • Estimate the percent reduction in number of shocks delivered per patient for treating spontaneous FVT episodes
  • Evaluate different possible predictors of ATP success: VT rate, underlying disease, Anti-Arrhythmic Drugs (AAD), infarct zone, etc.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Single (Participant)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • ICD indications (Class I-II A) according to the guidelines (patient with coronary artery disease [CAD] or non-CAD in primary or secondary ICD prevention)
  • Patients have been implanted with a Medtronic Marquis family ICD capable of ATP for FVT via VF

排除标准

  • Patient's life expectancy less than 1 year due to a non-cardiac chronic disease
  • Patient on heart transplant list which is expected in < 1 year
  • Patient's age less than 18 years
  • ICD replacements and upgrading (single chamber [SC] ICD® dual chamber [DC] ICD)
  • Unwillingness or inability to provide written informed consent
  • Enrollment in, or intention to participate in, another clinical study during the course of this study
  • Inaccessibility for follow-up at the study center
  • Ventricular tachyarrhythmias associated with reversible causes
  • Brugada syndrome, long QT and hypertrophic cardiomyopathy (HCM) patients
  • Other electrical implantable devices (neurostimulators, etc.)
  • Mechanical tricuspid valve

结局指标

主要结局

Efficacy of Anti Tachycardia Pacing (ATP) Therapy to Terminate Fast Ventricular Tachycardia (With Cycle Length of 240ms-320msec)

时间窗: one year

Termination of Ventricular Tachycardia is calculated as percentage of successfully terminated episodes, adjusted for multiple events with (GEE) method. This technique yields an average therapy efficacy and 95% CI that is based on number of patients, number of episodes per patient, and magnitude of the correlation between responses within patients.

次要结局

  • Efficacy of ATP in Successfully Treating FVT for Patients in Primary and Secondary Prevention(one year)
  • Acceleration Rate or Degenerated Into VF of ATP for Treating FVT in the 2 Arms(one year)
  • Percent Reduction in Shocks Delivered Per Patient for Treating FVT(one year)
  • Compare Likelihood of Syncopal Events Associated With FVT(one year)
  • Evaluate Different Possible Predictors of ATP Success(one year)