ADVANCE CRT - D: ATP Delivery for Painless ICD Therapy

Medtronic Cardiac Rhythm and Heart Failure2 个研究点 分布在 1 个国家目标入组 526 人开始时间: 2004年2月1日最近更新:
适应症

试验速览

阶段
4 期
状态
已完成
发起方
入组人数
526
试验地点
2
主要终点
Efficacy of Anti Tachycardia Pacing (ATP) Therapy (Burst, 8 Pulses, 88 %, 1 Sequence) to Terminate All Types of Ventricular Tachycardia.

研究概览

简要总结

To compare the efficacy of RV and BiV ATP for the termination of ventricular arrhythmias in patients who are candidates to a cardiac resynchronisation therapy (CRT) and have a Class I or IIA indication for ICD implantation.

The hypothesis of delivering ATP from different sites (RV or BIV) has never been evaluated in a prospective, controlled and randomized study.

详细描述

Main objective: Compare efficacy of ATP therapy (Burst, 8 pulses, 88 %, 1 sequence) to terminate all types of ventricular tachycardia (all VTs (FVT+VT)) when delivered in the right ventricle (RV) only versus both ventricles (BiV) resulting in a 10 % difference in favour of BIV ATP

Secondary objectives:

  • Compare efficacy of the first BiV and RV ATP (Burst, 8 pulses, 88 %) to terminate fast ventricular tachycardia (FVT)
  • Compare efficacy of the first BiV and RV ATP (Burst, 8 pulses, 88 %) to terminate slow ventricular tachycardia (slow VT)
  • Compare efficacy of BiV and RV ATP (all ATP therapies) to terminate slow ventricular tachycardia (slow VT)
  • Determine the rate of both FVT and VT episodes which are accelerated or degenerates into VF

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Single (Participant)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • CRT + ICD indications (Class I-IIA) according to the guidelines
  • Patients have been implanted with a Medtronic Marquis Family ICD capable of RV-ATP or BIV-ATP
  • Patients in chronic AF who undergo a complete AV ablation and that the complete AV block is confirmed at PHD

排除标准

  • Patient's life expectancy less than 1 year due to a non cardiac chronic disease
  • Patient on heart transplant list which is expected in < 1 year
  • Patient's age less than 18 years
  • Replacements and upgrades
  • Epicardial lead
  • Mechanical tricuspid valve
  • Ventricular Tachyarrhythmias associated with reversible causes
  • Unwillingness or inability to provide written informed consent
  • Enrollment in, or intention to participate in, another clinical study during the course of this study
  • Inaccessibility for follow-up at the study center

结局指标

主要结局

Efficacy of Anti Tachycardia Pacing (ATP) Therapy (Burst, 8 Pulses, 88 %, 1 Sequence) to Terminate All Types of Ventricular Tachycardia.

时间窗: one year

Termination of Ventricular Tachycardia is calculated as percentage of successfully terminated episodes, adjusted for multiple events with GEE method. This technique yields an average therapy efficacy and 95% CI that is based on number of patients, number of episodes per patient, and magnitude of the correlation between responses within patients.

次要结局

  • Compare Efficacy of the First BiV and RV ATP to Terminate FVT(one year)
  • Compare Efficacy of the First BiV and RV ATP to Terminate Slow VT(one year)
  • Compare Efficacy of BiV and RV ATP (All ATP Therapies) to Terminate Slow VT(one year)
  • Determine the Rate of Both FVT and VT Episodes Which Are Accelerated or Degenerates Into VF(one year)

研究者

发起方
Medtronic Cardiac Rhythm and Heart Failure
申办方类型
Industry

研究点 (2)

Loading locations...

相似试验

ADVANCE CRT - D: Antitachycardia Pacing (ATP)... | 临床试验