Validation of Wrist Type Oscillometric Blood Pressure Monitorings
试验速览
- 阶段
- 不适用
- 入组人数
- 100
- 主要终点
- Validation
研究概览
简要总结
Since the wrist devices have the advantage of being smaller, lighter, and more inexpensive than the upper arm devices, wrist devices become increasingly popular. However, it is of utmost importance to ensure that the BP measuring device can measure BP accurately. This study will be aimed to validate a wrist type oscillometric blood pressure monitoring with reference to the brachial blood pressure measured by auscultatory method with a mercury sphygmomanometer.
详细描述
High blood pressure (BP) is one of the major cardiovascular risk factors in adults. Home blood pressure monitoring is superior to office-based measurements for prediction of the future development of cardiovascular diseases and is recommended by major guidelines. To monitor BP at home, patients can use devices that measure BP either at the arm or at the wrist.
Among BP monitors, the wrist device is relatively easy to use, as the cuff can be used with one hand easily and correctly around the wrist than around the arm, and undressing is not necessary. Since the wrist devices have the advantage of being smaller, lighter, and more inexpensive than the upper arm devices, wrist devices become increasingly popular. However, it is of utmost importance to ensure that the BP measuring device can measure BP accurately. To make sure the performance of these devices, several guidelines have proposed protocols to validate these devices. This study will be aimed to validate a wrist type oscillometric blood pressure monitoring with reference to the brachial blood pressure measured by auscultatory method with a mercury sphygmomanometer.
研究设计
- 研究类型
- Observational
- 观察模型
- Other
- 时间视角
- Cross Sectional
入排标准
- 年龄范围
- 12 Years 至 —(Child, Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 是
入选标准
- •above 12 years old
排除标准
- •unclear Korotkoff sounds
- •wrist circumference outside of the designated range (13.5-23 cm)
- •arrhythmias
结局指标
主要结局
Validation
时间窗: 7 months
The present study will be aimed to evaluate the accuracy of the wrist type oscillometric blood pressure monitorings.
次要结局
未报告次要终点