Sample Type Validation: Finger Prick and Plasma vs Whole Blood in a Clinical Setting for the SpinChip High-sensitivity Cardiac Troponin I (SpinChip Hs-cTnI) Test

SpinChip Diagnostics ASA1 个研究点 分布在 1 个国家目标入组 180 人开始时间: 2024年4月12日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
180
试验地点
1
主要终点
Sample type validation

研究概览

简要总结

During a heart attack, the protein troponin I is released from the heart muscle into the bloodstream. Measurements of cardiac troponin in blood are used as an aid in the diagnosis of heart attack. The SpinChip hs-cTnI test is a new high-sensitive test for measuring the amount of cardiac troponin I in the bloodstream as an aid in the diagnosis of heart attack.

The purpose of this study is to demonstrate that different types of blood samples (finger prick, venous whole blood and plasma) return comparable results when analysed using the SpinChip hs-cTnI test. Blood samples from at least 150 patients will be analyzed and the testing will be carried out by healthcare personnel.

详细描述

Cardiac troponins are widely used as a biomarker to aid in the diagnosis of Acute myocardial infarction (AMI). These structural proteins are essential in regulating contraction in cardiac muscle cells, and they are sensitive and specific biochemical markers of myocardial damage.

During a heart attack, cardiac muscle cells are injured and release the cardiac marker troponin I (cTnI) into the bloodstream. High-sensitive troponin tests may detect the increase of cardiac troponin in blood within hours after the symptoms of a heart attack has started.

The SpinChip high-sensitivity cardiac troponin I (hs-cTnI) test is a new high-sensitive test for measuring troponin I in blood samples, and the analysis may be performed close to the patient (near-patient test). The results may be obtained within 10 minutes, compared to approximately 1 hour for normal laboratory analysis.

The SpinChip Platform consists of the SpinChip hs-cTnI test and the SpinChip Analyzer and may be used at the emergency department to evaluate patients presenting with symptoms of acute myocardial infarction (chest pain). The test can use blood from finger prick or venous blood samples, either as whole blood or separated into plasma.

The purpose of this study is to demonstrate that different sample types (finger prick, venous whole blood or plasma) give comparable results when analysed using the SpinChip hs-cTnI test when the testing is performed by healthcare professionals. Samples from minimum 150 patients with troponin levels covering the measuring range of the SpinChip hs-cTnI test will be collected and analysed.

研究设计

研究类型
Observational
观察模型
Other
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Able and willing to provide signed written informed consent
  • Subjects >18 years old
  • Clinical cardiac Troponin T concentration available

排除标准

  • Cognitive impairment precluding informed consent
  • Self-reported pregnancy
  • General clinical condition which affects the patient in a major way and the patient is considered to be in an unstable clinical condition
  • Previously included in the study with a sample within the measuring range of the SpinChip hs-cTnI test

结局指标

主要结局

Sample type validation

时间窗: 1 day

Establish if the hs-cTnI results obtained from finger prick capillary blood and fresh plasma are equivalent to results obtained for venous whole blood for the SpinChip hs-cTnI test when the analysis is performed by healthcare professionals. Equivalence between sample types will be evaluated using Bland Altman difference plots for evaluation of bias and regression analysis.

次要结局

  • Incidence of AEs, ADEs and DDs(1 day)