Investigation of the accuracy of the cardiac output measurement using two different minimally invasive cardiac output monitoring system - Investigation of the accuracy of the cardiac output measurement using two different minimally invasive cardiac output monitoring system

Tohoku University Hospital department of anesthesiology0 个研究点目标入组 29 人开始时间: 2016年7月31日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
29

研究概览

简要总结

Twenty-nine patients were enrolled. Compared with CITD, CIFT and CILR had a percentage error (PE) of 39.9% and 51.2%, respectively. The accuracy of CIFT or CILR is not clinically acceptable. Fourth-generation Flotrac was unreliable in high-SVRI. LiDCOrapid was inaccurate across a broad range of SVRI and minimally affected by SVRI.

研究设计

研究类型
Observational

入排标准

年龄范围
20years-old 至 ot applicable(—)
性别
All

入选标准

  • 未提供

排除标准

  • withdrawal of consent moderate or more aortic valve stenosis or regurgitation intra aortic balloon pumping extracorporeal membrane system left ventricular assist device thoracic aorta placement moderate or more tricuspid regurgitation low cardiac output