JPRN-UMIN000023313已完成未知
Investigation of the accuracy of the cardiac output measurement using two different minimally invasive cardiac output monitoring system - Investigation of the accuracy of the cardiac output measurement using two different minimally invasive cardiac output monitoring system
Tohoku University Hospital department of anesthesiology0 个研究点目标入组 29 人开始时间: 2016年7月31日最近更新:
适应症
试验速览
- 阶段
- 未知
- 状态
- 已完成
- 发起方
- 入组人数
- 29
研究概览
简要总结
Twenty-nine patients were enrolled. Compared with CITD, CIFT and CILR had a percentage error (PE) of 39.9% and 51.2%, respectively. The accuracy of CIFT or CILR is not clinically acceptable. Fourth-generation Flotrac was unreliable in high-SVRI. LiDCOrapid was inaccurate across a broad range of SVRI and minimally affected by SVRI.
研究设计
- 研究类型
- Observational
入排标准
- 年龄范围
- 20years-old 至 ot applicable(—)
- 性别
- All
入选标准
- 未提供
排除标准
- •withdrawal of consent moderate or more aortic valve stenosis or regurgitation intra aortic balloon pumping extracorporeal membrane system left ventricular assist device thoracic aorta placement moderate or more tricuspid regurgitation low cardiac output
研究者
相似试验
已完成
不适用
Cardiac Output by Nine Different Pulse Contour AlgorithmsCardiac OutputNCT02438228University Hospital Schleswig-Holstein30
已完成
不适用
The Precision of Pulmonary Artery Cardiac Output-measurements in Spontaneously Breathing PatientsHeart DiseasesNCT01944254Norwegian University of Science and Technology18
已完成
不适用
Comparing Cardiac Output Measurements in Critically Ill Children Using USCOM and Transthoracic ECHOCritical IllnessPediatricsNCT01460329University of North Carolina, Chapel Hill22
终止
不适用
Study of Echocardiogram Accuracy in Patients With Mitral Valve CalcificationMitral Annulus CalcificationNCT01442467Creighton University4
已完成
不适用
Accuracy of non-invasive cardiac output measurement in awake patientsPostoperative patients transferred to the intensive care unit under intubationJPRN-UMIN000044712agoya University40