MedPath

The efficacy of KneeAlign2 navigation system for Vanguard total knee arthroplasty -randomized controlled trial

Not Applicable
Conditions
Osteoarthritis of the knee joint
Registration Number
JPRN-UMIN000021436
Lead Sponsor
Osaka City University Graduate School of Medicine
Brief Summary

Six patients in the navigation group and 15 patients in the conventional group had an alignment which was > 2 degrees away from the neutral mechanical axis of tibia (p <0.05). The mean deviation from neutral alignment of tibia was 1.2 +/- 1.2 degrees (mean +/- SD) in the navigation group and 1.7 +/- 1.1 degrees in the conventional group (p <0.05). The mean deviation from neutral alignment of femur was 1.4 +/- 1.3 degrees in the navigation group and 1.9 +/- 1.4 degrees in the conventional group (p <0.05).

Detailed Description

Not available

Recruitment & Eligibility

Status
Complete: follow-up complete
Sex
All
Target Recruitment
100
Inclusion Criteria

Not provided

Exclusion Criteria

Other than osteoarthritis Infection Valgus knee

Study & Design

Study Type
Interventional
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Postoperative implant alignment (6 months postoperatively)
Secondary Outcome Measures
NameTimeMethod
Knee Society score, Oxford knee score, EQ5D (6 months postoperatively)
© Copyright 2025. All Rights Reserved by MedPath