Evaluating an Under-mattress Sleep Monitor Compared to a Peripheral Arterial Tonometry Device in the Diagnosis of Obstructive Sleep Apnea
- Conditions
- Sleep Apnea, Obstructive
- Interventions
- Diagnostic Test: Under-mattress sleep monitor
- Registration Number
- NCT04778748
- Lead Sponsor
- Mayo Clinic
- Brief Summary
A device has been developed which is placed under the mattress, thereby escaping contact with patient, which enables monitoring of sleep activity and has been shown in some studies to provide an estimate of the frequency of sleep disordered breathing. If the estimates of sleep disordered breathing from this new device are medically equivalent to current excepted home sleep apnea test devices within clinically important ranges of sleep disordered breathing, it may have several benefits. This study is designed to help answer several important clinical questions, namely, is the new device clinically equivalent to an established HSAT device for patients who have at least moderately severe OSA, and what is the effect of multiple nights on classification of severity and presence of OSA.
- Detailed Description
The design will be a prospective test of two diagnostic testing strategies. The tests will take place in the same patient concurrently, so the patient's results from one test will serve as the comparator with the other test device. No randomization or blinding is needed.
Patients will be recruited from the Mayo Clinic Center for Sleep Medicine (and another sleep center with enriched racial variety). Consecutive patients in whom the clinician has ordered a WatchPAT™ will be potential candidates for an opportunity to participate in the study. Patients will be made aware of the trial by invitation from members of the healthcare team. Interested patients will be evaluated according to the inclusion and exclusion criteria.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 60
Clinician suspects of having OSA due to presence of any:
- STOP-BANG ≥ 3,
- overnight oximetry with ODI ≥ 5,
- STOP-BANG≥ 2 and one or more of: disrupted sleep, cognitive complaints, paroxysmal or prior atrial-fibrillation (now in NSR), stroke, TIA
- Has a smartphone capable of running both Withings and Itamar's app
- Stable sleeping quarters and schedule (i.e., no overnight shifts) for at least 7 days
- Domicile has capable wireless internet service
- Uses short-acting nitrates within 3 hours of the study
- Has a permanent pacemaker
- Atrial fibrillation or sustained supraventricular arrhythmias
- Known congestive heart failure Class ≥ 1, or known LVEF< 45%
- Sustained hypoxemia or hypoventilation (results of RA daytime ABG show SaO2<90% or PaCO2>45 mmHg, or overnight oximetry shows mean SaO2<90%.)
- Advanced pulmonary disease (COPD GOLD stage ≥ II , pulmonary fibrosis with GAP ≥ 1 or significant dyspnea on exertion
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Arm && Interventions
Group Intervention Description Obstructive Sleep Apnea Under-mattress sleep monitor Participants will be provided with a Withings sleep monitor and instructed to begin using it. Effective data collection via the Withings unit should be confirmed for at least 7 consecutive nights, and the WatchPAT device will be used during one of those same 7 nights.
- Primary Outcome Measures
Name Time Method The apnea hypopnea index (AHI) 12 months The apnea hypopnea index (AHI) provided by one-night of testing with the WatchPAT™ is equivalent to the AHI provided by the mean of 7 nights use of the Withings™ sleep monitor. The equivalence margin will be an AHI≤5/hr.
- Secondary Outcome Measures
Name Time Method Sensitivity 12 months The Withings sleep monitor will have a sensitivity ≥ 0.825 to identify patients with moderate to severe OSA as assessed by a WatchPAT AHI≥15.
Trial Locations
- Locations (1)
Mayo Clinic in Rochester
🇺🇸Rochester, Minnesota, United States