Electrical Impedance Tomography (EIT) Monitoring: A Pilot Comparison to Standard of Care Assessments in Adults With Acute Lung Injury (ALI) or Acute Respiratory Distress Syndrome (ARDS)
Trial Snapshot
- Phase
- Not Applicable
- Status
- Completed
- Enrollment
- 13
- Locations
- 1
- Primary Endpoint
- Feasibility of EIT Monitoring in This Population of ARDS/ALI Patients
Study Overview
Brief Summary
Electrical impedance tomography (EIT) monitoring has been researched as a method to determine the spatial impedance distribution in a body cross section, but has yet to become an established clinical tool. EIT monitoring gives a dynamic, breath-to-breath measurement of both global and regional ventilation. Recently, there has been evidence that EIT monitoring has great potential to become a non-invasive bedside tool for assessment of regional lung ventilation without documented hazards. Potential applications include any adult patients in acute respiratory failure. Data collected from this research may contribute to improved patient safety outcomes.
PURPOSE:
The purpose of this pilot study is to examine the feasibility of using the EIT monitor in intensive care unit (ICU) setting on patients with acute respiratory failure and to compare the EIT monitor data to standard of care patient assessments. It is hypothesized that the EIT monitor, when applied to adults in acute respiratory failure, will correlate with conventional standard of care assessments for these patients.
Detailed Description
INTRODUCTION:
Electrical impedance tomography (EIT) monitoring has been researched as a method to determine the spatial impedance distribution in a body cross section, but has yet to become an established clinical tool. EIT monitoring gives a dynamic, breath-to-breath measurement of both global and regional ventilation. Recently, there has been evidence that EIT monitoring has great potential to become a non-invasive bedside tool for assessment of regional lung ventilation without documented hazards. Potential applications include any adult patients in acute respiratory failure. Data collected from this research may contribute to improved patient safety outcomes.
PURPOSE:
The purpose of this pilot study is to examine the feasibility of using the EIT monitor in ICU setting on patients with acute respiratory failure and to compare the EIT monitor data to standard of care patient assessments. It is hypothesized that the EIT monitor, when applied to adults in acute respiratory failure, will correlate with conventional standard of care assessments for these patients.
METHOD:
Study Design
- Study Type
- Interventional
- Allocation
- Na
- Intervention Model
- Single Group
- Primary Purpose
- Basic Science
- Masking
- None
Eligibility Criteria
- Ages
- 18 Years to — (Adult, Older Adult)
- Sex
- All
- Accepts Healthy Volunteers
- No
Inclusion Criteria
- •Patients will be included in this study if the following criteria have been met:
- •Males and females > 18 years old (non-pregnant, non-lactating females).
- •Partial pressure of arterial oxygen (PaO2) / Fraction of inspired oxygen (FiO2) ratio < 300 torr.
- •Signed informed consent has been obtained.
Exclusion Criteria
- •Females who are pregnant or lactating.
- •Patients with skin abrasions or wounds in the chest area.
- •Patients receiving dialysis.
- •Patients with pacemakers or paced EKG rhythms.
Arms & Interventions
Adults with ARDS or ALI
Adults with PaO2/FiO2 ratio less than 300. Consented patients will be placed on EIT monitor. The only intervention is the addition of Electrical Impedance Tomography monitoring using the Chest belt with 16 electrodes connected to the EIT device. No changes to standard patient care will occur other than collecting EIT monitor data.
Intervention: Electrical Impedance Tomography monitoring (Device)
Outcomes
Primary Outcomes
Feasibility of EIT Monitoring in This Population of ARDS/ALI Patients
Time Frame: At the start of monitoring once the patient was consented and enrolled.
Feasibility for the purposes of our study was the ability to apply the device to a diverse population of ARDS/ALI patients and obtain EIT data from the device.
Secondary Outcomes