Prospective Study of Micro-CT Imaging to Quantify Dialyzer Clotting in Patients With End Stage Kidney Disease Undergoing Hemodialysis

University Hospital, Ghent2 个研究点 分布在 1 个国家目标入组 10 人开始时间: 2024年1月8日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
10
试验地点
2
主要终点
Intrapatient variability of dialyzer fiber patency

研究概览

简要总结

Different cross-over studies have been performed investigating dialyzer fiber patency in different dialysis setups. Herewith, post dialysis micro computed tomography (CT) images of the dialyzer were compared. For the best interpretation of such results, one should have an idea about the intrapatient variability. There is also no clue about the impact of long distance transportation and long cold storage on the reproducibility of the micro CT images. Another bottle neck is that, up till now, no biochemical parameter or test has been found associated with the outcome of dialyzer fiber patency post dialysis.

The present study therefore aims at determining the intrapatient variability and the impact on the micro CT results of long distance transportation and long cold storage of the dialyzers. Also, whole blood thrombin generation tests are performed to look for associations with the micro CT results.

详细描述

During recent years, different coagulation studies were performed using micro computed tomography (CT) scanning of hemodialyzers post dialysis as a marker for fiber blocking. Each of these studies investigated the impact of either anticoagulation strategy, dialysis strategy, or hemodialyzer type in a cross-over study design in which each patient served as his/her own control.

To date, we still need to answer some remaining questions related to studies using micro CT before we can expand the research and clinical applications to a larger scale:

First, the power of such cross-over studies, with a single session per study arm, depends also partially on the variability over time in the micro CT results within a single patient. While the resolution (25µm) and reproducibility of the micro CT process is established, patients' coagulation status will cause the largest variability in dialyzer outcome among consecutive dialysis sessions, e.g. variation due to infection or vascular access problems. This type of intrapatient variability, quantifying dialyzer outcome with micro CT from comparable but different dialysis sessions, has not yet been studied. Evaluation of the inter- and intra-patient variability of micro CT of dialyzers in chronic hemodialysis patients will inform whether this novel method can be used as an endpoint in future interventional trials of novel anticoagulants.

Second, the micro CT imaging technique, recently paraphrased as the gold standard to quantify clotting, is a very valuable technique in research settings. Clinical practice would profit more from a less burdensome marker of hemodialyzer performance as measured during the dialysis session (not only at the end of a dialysis session). However, none of the available online dialysis machine parameters (i.e., transmembrane pressure, arterial and venous pressure, online clearance monitoring), nor visual scoring of the dialyzer or venous chamber, nor assessment of dialyzer mass post-treatment correlated well with the results of micro CT. Also, the commonly used biochemical parameters only focus on one aspect in the coagulation cascade or/and are not sensitive enough to measure variations in the coagulation system of the patient. As a consequence, they are poorly related to dialyzer outcome as quantified with micro CT. While thrombin generation (TG) is commonly determined in plasma to identify global coagulation phenotype, whole blood TG (WB-TG) tests better mimic physiology by involving also the intrinsic blood cells and platelets, making it a potential biomarker test for coagulation, and this requires investigation.

Third, the previously described studies were all executed at the Ghent University Hospital where dialyzers are prepared for scanning on the spot (i.e., rinsed, dried and stored at 5°C), and with only a short transport time (15min) from the clinic to the micro CT scanner. No hard evidence however exists whether dialyzers, after being prepared for micro CT in a different place, might be transported by flight in e.g. isolated boxes (lasting hours to days). This study will be conducted in collaboration with Regeneron (New York) where a subset of 10 dialyzers will be shipped, imaged using Regeneron's micro CT scanner, and shipped back to Ghent University for repeat imaging.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • stable chronic hemodialysis patient
  • well functioning vascular access

排除标准

  • ultrafiltration rate higher than 4 litre per session
  • use of antiplatelets or anticoagulants (apart from acetylsalicylic acid)
  • known coagulation disorder
  • active inflammation
  • malignancy

结局指标

主要结局

Intrapatient variability of dialyzer fiber patency

时间窗: 2 months

The intrapatient variability of the dialyzer fiber patency (i.e. percentage open fibers post dialysis) and the intraclass correlation coefficient (ICC) of fiber patency are calculated from the results of the post dialysis micro CT images of the dialyzer, as scanned after three consecutive dialysis sessions with the same anticoagulation dosing: one week with full anticoagulation and one week with reduced (1/4) anticoagulation. Outcome = intrapatient variability of the percentage open fibers. Tool = micro CT scanning and computer-based determination of fiber patency.

次要结局

  • Impact of dialyzer storage and transport on micro CT images of dialyzer fibers(3 months)
  • Association between whole blood thrombin generation test and fiber patency(2 months)
  • Difference in fiber patency using normal versus reduced anticoagulation dosing.(2 months)