Comparison of three commonly used supraglottic devices for their efficacy
- Conditions
- undergoing elective surgery
- Registration Number
- CTRI/2017/08/009534
- Lead Sponsor
- Department of anesthesia
- Brief Summary
Our study was conducted to compare proseal laryngeal mask airway, I-gel and supreme laryngeal mask airway. All three devices were easy to insert,had 100% 1st attempt success rate,elicited similar hemodynamic response and allowed easy passage of nasogastric tube. Proseal LMA provided highest oropharyngeal leak pressure ( 32.64 cmH2O ), LMA Supreme provided higher oropharyngeal leak pressure compared to i-gel.
I-gel And Supreme LMA were quick to insert (13.50 sec and 14.50 sec.All three devices are reliable airway devices,producing efficient airway leak pressures with a very low incidence of trauma and airway morbidity,no incidence of regurgitation and aspiration in any of the patients.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- Completed
- Sex
- All
- Target Recruitment
- 84
ASA Grade 1 OR 2.
Known or predicted difficult airway ,full stomach or increased risk of aspiration (GERD, pregnancy, hiatus hernia).
Study & Design
- Study Type
- Interventional
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Oropharyngeal leak pressure 1 min, 15 min
- Secondary Outcome Measures
Name Time Method time of insertion,ease of insertion, hemodynamic response and morbidity associated till 30 min of surgery
Trial Locations
- Locations (1)
Maulana Azad medical college
🇮🇳Central, DELHI, India
Maulana Azad medical college🇮🇳Central, DELHI, IndiaAnisha SinghPrincipal investigator9810341603anisha_sngh@yahoo.co.in