Randomised Comparison of the Octaray and Pentaray Catheters

Not Applicable
Not yet recruiting
Conditions
Atrial Fibrillation
Interventions
Other: Left atrial mapping
Registration Number
NCT05988424
Lead Sponsor
University Hospital Southampton NHS Foundation Trust
Brief Summary

In this study we aim to compare the mapping performance in the left atrium of the established Pentaray catheter to the newer Octaray catheter.

Detailed Description

Compared with conventional point by point mapping, the use of a multipolar high density Pentaray catheter (Biosense Webster, Diamond Bar, CA, USA) has been found to produce more rapid, accurate maps. This catheter has 20, 2mm2 electrodes. Newer, high density mapping catheters have also been introduced, capable of collecting detailed maps rapidly. Such catheters include the Orion catheter (Boston Scientific, Marlborough, MA, USA) which has 64, 0.4mm2 electrodes and has been shown to be safe and effective in real world usage, and the HD Grid (Abbott Cardiovascular, Lake Forest, IL, USA) which has 16, 1mm electrodes and has been shown to be associated with shorter procedure and fluoroscopy times and better success rates than conventional circular mapping catheters when used for AF ablation.

Of the above high density catheters, the Pentaray is integrated into the Carto3 system. An advantage of this platform for mapping is the ability to use the Tissue Proximity Indicator (TPI) Filter to reduce the number of points collected without contact and improve the accuracy of the subsequent voltage map.

A novel multipolar catheter, iterating on the previous Pentaray catheter is now being introduced: Octaray. This has 48, 0.9mm2 electrodes arranged on 8 splines. In preclinical studies, Octaray has been found to be collect maps faster, at higher density and with greater accuracy for conduction gaps than the Pentaray. The new catheter has been used safely in clinical cases.

Methods This is a randomised, prospective, single centre, interventional study, Thirty patients listed for redo AF ablation procedures on clinical grounds will be included in the study. Procedures will be under conscious sedation or general anaesthetic as per the preference of the operator, as will the mode of transeptal access and sheaths used to deliver the mapping catheter to the left atrium.

At the start of the case, two consecutive LA maps will be taken by the same operator using the Pentaray and Octaray catheters. In both cases, throughout the study the same catheter types will be used. For the Pentaray catheter, this will be the standard 4-4-4 spaced version. The Octaray for the study will be the larger, 20mm spline version, with the 3-3-3-3-3 spacing.

To minimise any learning effect from repeated map collections in the same chamber, patients will be randomly assigned, using block randomisation, as to whether their first LA map will be collected using the Octaray or Pentaray. The randomised group the patient is in will be kept in a sealed envelope which will only be opened on the day of the procedure. All maps will be respiratory gated. The same sheath will be used for both catheters. For the study protocol, maps will be taken with first one mapping catheter and then the alternative. The rhythm during collection will be the same for both maps. Ideally patients would be in sinus rhythm, with maps collected with coronary sinus pacing. If there were challenges maintaining sinus rhythm during mapping, maps can be taken in AF (though not a mixture of rhythms). The setup in terms of window of interest and references will also be identical. The map detail level will be left at 16 for all maps. The map maximum point density will be left as nominal. The fill and colour thresholds will be kept unchanged at nominal values. No ablation will be permitted between the two maps being collected. When the second map is being collected using the alternative catheter, this will be a new map and not a remap and the prior map will be hidden from the operator.

In order to control for any matrix collected by the initial catheter in taking the first map, a completely new study will be used for the second map.

TPI will be on for the entire collection period for both maps. The time when map collection was started, as well as the fluoroscopy time at that point will be noted. These values will also be noted at the end of the map collection. The aim of map collection will be to get entire coverage of the geometry with voltage data. Where this was not possible will be noted. The time for the catheter to start collecting voltage data will also be recorded, to judge the delay in TPI filter accepted points for the two catheters.

LA Map volumes will be collected from Carto3. The pulmonary veins and any left ventricle inadvertently collected will be removed from the geometries to allow quantitative assessment.

Once the mapping phase has been completed, the operator would complete the procedure as to their usual practice.

There will be no follow up needed for the study. Any acute complications will be noted - acute in this case being defined as complications occurring up to 30 days post-procedure.

Recruitment & Eligibility

Status
NOT_YET_RECRUITING
Sex
All
Target Recruitment
30
Inclusion Criteria

Patients >18 years listed for redo AF ablation on clinical grounds GA or LA/sedation cases

Exclusion Criteria

Age <18 years Pregnancy Unable or unwilling to consent Mechanical mitral valve Replacement Enrolled in another clinical study that could confound the results of this study (note: patients enrolled in complementary study are eligible for enrolment)

Study & Design

Study Type
INTERVENTIONAL
Study Design
CROSSOVER
Arm && Interventions
GroupInterventionDescription
OctarayLeft atrial mappingInitial map collected with Octaray
PentarayLeft atrial mappingInitial map collected with Pentaray
Primary Outcome Measures
NameTimeMethod
Time to collect left atrial geometry1 year

Time to collect left atrial geometry

Secondary Outcome Measures
NameTimeMethod
Map point density1 year

Map point density

Time to start point collection, based on fulfilment of TPI criteria1 year

Time to start point collection, based on fulfilment of TPI criteria

Low voltage area comparison1 year

Low voltage area comparison

Number and size of PVI gaps identified1 year

Number and size of PVI gaps identified

Volume of LA geometry1 year

Volume of LA geometry