Comparison of SureCALL® EMG Labor Monitor® and Tocodynamometer to Intrauterine Pressure Catheter Measurement in Pregnant Women

Reproductive Research Technologies, LP3 个研究点 分布在 1 个国家目标入组 20 人开始时间: 2008年8月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
发起方
入组人数
20
试验地点
3
主要终点
Comparison of SureCALL® and TOCO Detection of Contraction Events, as Compared to the IUPC

研究概览

简要总结

This measurement study was designed to evaluate the performance of SureCALL® EMG Labor Monitor® (SureCALL®) to the predicate Tocodynamometer device (TOCO) using the performance of an Intrauterine Pressure Catheter device (IUPC) as the "gold standard".

详细描述

The SureCALL® EMG Labor Monitor® (SureCALL®) is a transabdominal electromyography monitor intended to measure intrapartum uterine activity. It is intended for use on term pregnant women who are in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The device is intended for use by healthcare professionals in a clinical setting.

This measurement study was designed to evaluate the performance of SureCALL® to the predicate Tocodynamometer device (TOCO) using the performance of an Intrauterine Pressure Catheter device (IUPC) as the "gold standard".

This study involved 20 women at term, in labor, at three clinical sites. Each study subject was instrumented with three technologies for measuring uterine activity:

  1. a tocodynamometer attached to the maternal abdomen
  2. a set of abdominal surface electrodes for uterine electromyography, and
  3. an intrauterine pressure catheter.

Using IUPC as the "gold standard", this study methodology allowed a three way comparison for evaluating how well the SureCALL® system performed compared to the TOCO technology.

研究设计

研究类型
Observational
观察模型
Case Only
时间视角
Cross Sectional

入排标准

年龄范围
15 Years 至 40 Years(Child, Adult)
性别
Female
接受健康志愿者

入选标准

  • Singleton Pregnancy
  • Indicated for IUPC
  • Gestational ages from 37 to 41 weeks
  • Informed consent required

排除标准

  • Multifetal Pregnancy
  • Not Indicated for IUPC
  • Gestation ages below 37 or above 41 weeks
  • Informed consent not given

结局指标

主要结局

Comparison of SureCALL® and TOCO Detection of Contraction Events, as Compared to the IUPC

时间窗: 9 - 42 Minutes

Contraction timing as measured by the SureCALL® and contraction timing as measured by the TOCO, both compared to the contraction timing as measured by the IUPC gold standard. The contraction timing values of SureCALL® and TOCO were then compared.