MedPath

K2 Medical KALPA X™ Mapping, Imaging and Navigation Device

Not Applicable
Terminated
Conditions
Left Atrial Appendage Aneurysm
Interventions
Device: KALPA X™
Registration Number
NCT05155943
Lead Sponsor
Cara Medical Ltd
Brief Summary

This prospective, multi-center, non-randomized, single-blinded, open-label, single-arm, comparator study will enroll up to 100 eligible subjects from multiple sites undergoing LAAC procedure to evaluate the safety and performance of the KALPA™ mapping, imaging and navigation device in patients undergoing Left Atrial Appendage Closure (LAAC).

Detailed Description

All subjects will undergo pre-procedural cardiac computed tomography angiography (CCTA) no more than 1 year prior to the performance of the LAAC procedure to evaluate left atrial appendage (LAA) anatomy. Before and after a transseptal puncture, the right atrium (RA), left atrial appendage (LAA), and left upper superior pulmonary vein (LUSPV) will be mapped using the KALPA™ system in combination with a commercially available catheter. Subjects will then undergo LAAC as per routine clinical care. Following implantation of the LAAC device, subjects will be assessed for the peri-device leak (PDL) with a transesophageal echocardiogram (TEE) as well as PDL assessment using the KALPA X™ system (the operator will be blinded to KALPA X™ PDL reading). The entire LAAC procedure will be conducted as customary practice using standard and approved off-the-shelf equipment (body surface electrodes, catheters, etc.) the operator will be blinded to the KALPA™ and KALPA X™ system information. The physician will neither use nor rely on any of the KALPA™ or KALPA X™system output for clinical decision-making.

All patients will then undergo cardiac computed tomography angiography (CCTA) and TEE at 45 days following the index LAAC procedure to assess residual PDL. Intra-procedure PDL post-LAAC assessed by KALPA X™ will be compared to the detection of PDL by TEE. The value of PDL as measured by TEE compared to KALPA X™ at the time of the index LAAC procedure in predicting PDL at 45 days post-LAAC as assessed by CCTA will be compared.

Recruitment & Eligibility

Status
TERMINATED
Sex
All
Target Recruitment
18
Inclusion Criteria
  1. The patient must be ≥ 18 years of age.
  2. The patient meets indications for LAAC procedure.
  3. LAAC will be performed using the AMPLATZER Amulet device (AMPLATZER™ Amulet™ LAA Occluder, Abbott Vascular, Santa Clara, CA, USA), or, WATCHMAN (WATCHMAN™ LAAC Device or WATCHMAN FLX™ LAAC Device, Boston Scientific Marlborough, MA, USA) devices, or LAmbre LAA closure system (Lifetech Scientific Co, Ltd, Shenzhen, China), or any other LAA closure device commercially available at the discretion of the principal investigator.
  4. The patient is willing to comply with specified follow-up evaluations.
  5. The patient, or legally authorized representative, has been informed of the nature of the study, agrees to its provisions and has been provided written informed consent, approved by the appropriate Medical Ethics Committee (EC)
Exclusion Criteria
  1. Pregnant or nursing subjects. Female subjects of child-bearing potential must have a negative pregnancy test done within 7 days prior to index procedure per site standard test.
  2. Patients with known atrial septal defect or atrial septal repair or closure device that would preclude transseptal puncture for LAAC.
  3. Patients who have a planned treatment with any other investigational device or procedure during the study period.
  4. Contraindications to transesophageal echocardiography (TEE), including recent esophageal or gastric surgery (within 6 months), esophageal obstruction (stricture or tumor), depressed respiratory function, active upper gastrointestinal bleeding, perforate viscous (known or suspected), or uncooperative patient.
  5. Contraindications to cardiac computed tomography angiography (CCTA) including pregnancy or hypersensitivity to iodinated contrast agent

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
KALPA XKALPA X™The KALPA™ and KALPA X™ systems will be used during the performance of the LAAC procedure that is performed using customary and conventional tools and imaging technologies as currently performed in the participating center.
Primary Outcome Measures
NameTimeMethod
LAA maximum diameter and perimeter assessed by KALPATM compared to cardiac CTUp to 24 hours

Measures of LAA maximum diameter and perimeter obtained by KALPATM performed during the LAAC procedure will be compared to these measurements obtained by the pre-procedure cardiac CT

Multi-Modality Assessment of Immediate Post-LAAC Peri-Device Leak (PDL) by TEE and KALPA X™Up to 24 hours

PDL measured by KALPA X™ (defined below) will be assessed compare to the PDL detected by TEE

Number of participants with KALPA X™ device-related adverse eventsUp to 24 hours

Safety of using KALPA X™ during a cardiac procedure, specifically the safety of using KALPA™ during the LAAC procedure, by collecting the records of device related AE's

Secondary Outcome Measures
NameTimeMethod
Follow-up PDLUp to 45 days

The predictive power of intra-procedural TEE and KALPA X™ for the detection of PDL will be compared to 45 day post-LAAC PDL as detected by TEE and CCTA

Trial Locations

Locations (2)

Policlinico San Donato

🇮🇹

Milano, Italy

Vilnius University Hospital Santaros Klinikos

🇱🇹

Vilnius, Lithuania

© Copyright 2025. All Rights Reserved by MedPath