Accuracy of Noninvasive Pulse Oximeter Sensor (MightySat RX)
- Conditions
- Healthy
- Interventions
- Device: MightySat RX Pulse Oximeter Sensor
- Registration Number
- NCT03125135
- Lead Sponsor
- Masimo Corporation
- Brief Summary
In this study, the level of oxygen within the blood will be reduced in a controlled manner by reducing the concentration of oxygen the study volunteer breathes. The accuracy of a noninvasive pulse oximeter sensor will be assessed by comparison to the oxygen saturation measurements from a laboratory blood gas analyzer.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 16
- Subjects must understand and consent to be in the study.
- American Society of Anesthesiology Class 1 (Healthy subjects without any systemic disease at all).
- Subjects who have any systemic disease at all.
- Subjects who do not understand the study and the risks.
- Smokers.
- Subjects who are pregnant.
- Subjects having either signs or history of peripheral ischemia.
- Others deemed ineligible by the clinical staff.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Test Subject MightySat RX Pulse Oximeter Sensor All subjects are enrolled into the test group and all subjects receive the MightySat RX Pulse Oximeter Sensor.
- Primary Outcome Measures
Name Time Method Accuracy of Sensor 1-5 hours Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square error (Arms) value.
Accuracy of Sensor during motion 1-5 hours Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample during motion conditions and calculating the arithmetic root mean square error (Arms) value.
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Masimo Corporation
🇺🇸Irvine, California, United States