Set up for in-Vivo Measurement of Dynamic Intraocular Pressure (IOP)
- Conditions
- Intraocular Pressure
- Interventions
- Device: phacoemulsifcation
- Registration Number
- NCT00856141
- Lead Sponsor
- Iladevi Cataract and IOL Research Center
- Brief Summary
In this modern era, surgeons prefer to perform emulsification through small micro incisions. To accomplish this safely, surgeons raise irrigation bottle heights or use forced infusion systems.We believe that this sustained rise in the IOP during micro-incision phacoemulsification for a substantial period could be detrimental to the ocular structures. Therefore, intra-operative IOP pertaining to micro-incision cataract surgery is an emerging concern.
Conventional systems for measuring IOP generate an isolated graph of continuous pressure recording. Interpretation of the IOP fluctuations from these graphs requires a two step superimposition approach to characterize the IOP changes.
- Detailed Description
The aim of the present study was to describe a new setup for intra-operative monitoring of IOP in a clinical setting and validate it by correlating IOP differences arising with different phacoemulsification parameters in eyes undergoing microincision coaxial phacoemulsification.
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 20
- patients undergoing phacoemulsification for uncomplicated senile cataracts
- Cataracts from grade 1 to 3
- presence of glaucoma,
- diabetic retinopathy,
- retinal vascular disease,
- high myopia (defined as AL > 25mm),
- uveitis,
- previous ocular trauma or surgery
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- PARALLEL
- Arm && Interventions
Group Intervention Description 1. High fluidic parameters phacoemulsifcation Bottle height varied from 90-110cms, fixed aspiration flow rate 40cc/min, vacuum upto 650mmHg depending on the grade of cataract 2. Low fluidic parameters phacoemulsifcation Bottle height varied from 70-90cms, fixed aspiration flow rate of 25cc/min, vacuum upto 400mmHg, depending on the grade of cataract
- Primary Outcome Measures
Name Time Method IOP fluctuation during each phase of surgery intra-operative
- Secondary Outcome Measures
Name Time Method comparison of IOP fluctuations between high and low parameters group intra-operative
Trial Locations
- Locations (1)
Raghudeep Eye Clinic
🇮🇳Ahmedabad, Gujarat, India