Tatsu/Tradewind System User Performance Evaluation
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 118
- 试验地点
- 2
- 主要终点
- Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
研究概览
简要总结
This clinical trial is designed to assess an investigational blood glucose monitoring system with fingerstick blood in the hands of untrained intended users.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Type 1 or type 2 diabetes
- •18 years of age and older
- •Able to speak, read, and understand English
- •Willing to complete all study procedures
排除标准
- •Previously participated in a study using this investigational system
- •Working for a medical laboratory, hospital, or other clinical setting that involves training on and clinical use of blood glucose monitors
- •Pregnancy
- •Hemophilia or any other bleeding disorder
- •Working for a competitive medical device company or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company
- •A condition which, in the opinion of the investigator, would put the person or study conduct at risk (reason for exclusion will be clearly documented by investigator or designee).
结局指标
主要结局
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method
时间窗: 1 hour
Untrained subjects with diabetes tested self-test fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS), which included an investigational meter and sensor. BGM results were compared with capillary plasm BG results obtained with a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG meter results were used to calculate the number of BG results within +/- 15 mg/dL (for reference BG results \<75mg/dL) or +/- 20% (for reference BG results \>=75mg/dL) of the reference method results (YSI capillary plasma).
次要结局
- Number of Study Staff Results Within +/- 15mg/dL (<75 mg/dL) or Within +/- 20% (>=75 mg/dL) of Laboratory Glucose Method When Testing Subject Blood Glucose (BG)(1 hour)