Performance of the TATSU/Tradewind Blood Glucose Monitoring System
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 96
- 试验地点
- 2
- 主要终点
- Percent of Venous Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
研究概览
简要总结
The purpose of this study is to evaluate the performance and acceptability of a new blood glucose monitoring system, which includes meter and sensor strip.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Diagnostic
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 —(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Type 1 or 2 diabetes
- •Age 18 years or older
- •Routinely performs blood glucose self testing at home, at least once per day
- •Able to speak, read, and understand English.
- •Is willing to complete all study procedures
排除标准
- •Currently pregnant
- •Hemophilia or any other bleeding disorder
- •Taking prescription anticoagulants or has clotting problems that may prolong bleeding. Taking aspirin daily is not excluded
- •Subject employee of competitive medical device company
- •Cognitive disorder or other condition, which in the opinion of the investigator, would put the person at risk or seriously compromise the integrity of the study
结局指标
主要结局
Percent of Venous Blood Glucose (BG) Results Within +/- 5 to 20 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
时间窗: 2 hours
Healthcare professionals (study staff) used an investigational blood glucose monitoring system (BGMS) with subject venous blood. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results \<75mg/dL) or +/- 5% to 20% (for reference BG results \>=75mg/dL) of the reference method results.
Percent of Capillary Blood Glucose (BG) Results Within +/- 5 to 20mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method
时间窗: 2 hours
Subjects with diabetes used an investigational blood glucose monitoring system (BGMS) with subject capillary blood. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 20 mg/dL (for reference BG results \<75mg/dL) or +/- 5 to 20% (for reference BG results \>=75mg/dL) of the reference method results.
次要结局
- Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension)(2 hours)