NCT03976986CompletedNot Applicable

Assessment of Portable Oxygen Concentrators in Infants Undergoing Hypoxic Challenge Testing. A Randomized Crossover Study.

Hospital Son Espases1 site in 1 country26 target enrollmentStarted: February 11, 2016Last updated:
Conditions
Interventions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Sponsor
Enrollment
26
Locations
1
Primary Endpoint
Assessment of portable oxygen concentrators (POCs) to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).

Study Overview

Brief Summary

Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment. Onboard oxygen administration remains controversial. The Federal Aviation Administration approved portable oxygen concentrators (POCs) for onboard oxygen supply but there is lack of evidence about the use, especially in children. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

Detailed Description

According to a forecast by International Air Transport Association the number of people travelling on commercial aircrafts is predicted to rise up to 8.2 billion passengers in 2037. Therefore evidence-based flight recommendations will gain importance in the future, especially for patients suffering from chronic diseases. Hypoxic Challenge Testing (HCT) is the recommended method for inflight hypoxia risk assessment where nitrogen is introduced in a plethysmograph reducing FiO2 (fraction of inspired oxygen) to 0.15. Oxygen supply is recommended if PaO2 (partial pressure of oxygen in arterial blood) drops <50-55 mmHg (in adults) or Sat.O2 (oxygen saturation) ≤ 85% (in infants) where non-invasive pulse oximetry is the recommended method for hypoxia assessment. Onboard oxygen administration remains controversial. The effectiveness of pulsed-dosed systems remains unclear in small patients due to shallow breathing patterns. The aim of our study is to establish the effectiveness and safety of POCs in infants undergoing HCT.

Study Design

Study Type
Interventional
Allocation
Randomized
Intervention Model
Crossover
Primary Purpose
Treatment
Masking
None

Eligibility Criteria

Ages
— to 16 Years (Child)
Sex
All
Accepts Healthy Volunteers
No

Inclusion Criteria

  • Patients with baseline Sat.O2 >94% AND
  • <1 year with neonatal respiratory disease
  • required oxygen supply in the last 6 months
  • chronic respiratory disease (eg cystic fibrosis, obstructive or restrictive pulmonary disease) with FEV 1 (forced expiratory volume in 1 second) or FVC (forced vital capacity) <50%.

Exclusion Criteria

  • acute respiratory infection

Arms & Interventions

cPOC - pPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by cPOC (continuous-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by pPOC (pulsed-flow).

Intervention: continuous-flow portable oxygen concentrator (cPOC) (Device)

cPOC - pPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by cPOC (continuous-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by pPOC (pulsed-flow).

Intervention: pulsed-flow portable oxygen concentrator (pPOC) (Device)

pPOC - cPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by pPOC (pulsed-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by cPOC (continuous-flow).

Intervention: continuous-flow portable oxygen concentrator (cPOC) (Device)

pPOC - cPOC

Experimental

Randomized crossover study (according to a random number table): Patients are allocated randomly to two study groups (cPOC/pPOC). In case of Sat.O2 drop ≤ 85% during HCT, oxygen is administered by pPOC (pulsed-flow). For patients who show a positive POC hypoxic reversal, HCT is repeated at 24 hours and oxygen is administered by cPOC (continuous-flow).

Intervention: pulsed-flow portable oxygen concentrator (pPOC) (Device)

Outcomes

Primary Outcomes

Assessment of portable oxygen concentrators (POCs) to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).

Time Frame: 20 minutes

Hypoxia (Sat.O2 drop ≤ 85%) measured by Masimo SET Radical-7 Electron® pulse oximeter is induced performing HCT. Thereafter oxygen is administered through cPOC or pPOC until baseline Sat.O2 (Sat.O2 \>93%) is achieved.

Secondary Outcomes

  • POC comparison to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)
  • Relate patient age (months) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)
  • Relate patient weight (kilograms) to POCs capacity to change HCT induced hypoxia (Sat.O2 drop ≤ 85%).(20 minutes)