Measurement of Expiratory Gas Temperature in Intubated Patients to Estimate Expiratory Absolute Humidity

Laval University1 个研究点 分布在 1 个国家目标入组 70 人开始时间: 2026年4月1日最近更新:
干预措施

试验速览

阶段
不适用
状态
尚未招募
入组人数
70
试验地点
1
主要终点
Expiratory gas temperature

研究概览

简要总结

Adequate humidification is essential for intubated patients because dry medical gases can cause thick secretions and dangerous endotracheal tube blockages. A clear correlation exists between exhaled and inhaled humidity delivered by heat and moisture exchanger. The objective of this study is to measure the temperature of exhaled gases in intubated patients in order to accurately calculate the absolute humidity during the expiratory phase. These data will allow for a more accurate reproduction of real clinical conditions during bench testing and will improve the assessment of heat and moisture exchanger performance device evaluation.

详细描述

Previous clinical studies suggest that real expiratory humidity in intubated patients is close to 35 mg H₂O/L, with expiratory temperatures around 32-33 °C or slightly lower. However, data are limited, heterogeneous, and often derived from anesthetized patients. Therefore, accurate measurement of expiratory temperature in intensive care patients under normothermia and across varying body temperatures is necessary to define realistic simulation standards and ensure reliable heat and moisture exchanger performance evaluation on a bench. A relationship between expiratory gases temperature and expiratory humidity exist.

研究设计

研究类型
Observational
观察模型
Cohort
时间视角
Prospective

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Adults (≥18 years)
  • Intubated and mechanically ventilated in the intensive care unite
  • Using passive humidification (Heat and Moisture exchanger)

排除标准

  • PEEP set < 12 cmH₂O
  • Inhaled nitric oxide therapy
  • Tracheostomy

研究组 & 干预措施

Normothermic patient

Normothermic patient under mechanical ventilation with passive humification (heat and moisture exchanger) with body temperature between 36.0 and 38.0 Celsius degrees

干预措施: Expiratory gases temperature (Other)

Hypothermic patient

Hypothermic patient under mechanical ventilation with passive humification (heat and moisture exchanger) with body temperature < 36.0 Celsius degrees

干预措施: Expiratory gases temperature (Other)

Hyperthermic patient

Hyperthermic patient under mechanical ventilation with passive humification (heat and moisture exchanger) with body temperature > 38.0 Celsius degrees

干预措施: Expiratory gases temperature (Other)

结局指标

主要结局

Expiratory gas temperature

时间窗: H0, up to 60 minutes after inclusion

Measure expiratory gas temperature immediately before the passive humidifier

Relationship between body temperature and expiratory temperature.

时间窗: H0, up to 60 minutes after inclusion

Evaluate the relationship between body temperature and expiratory temperature

次要结局

  • Estimate expiratory absolute humidity based on measured temperature(H0, up to 60 minutes after inclusion)

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

François Lellouche

Principal Investigator

Laval University