A prospective clinical study of venous arterial blood manipulation system on prevention with arterial line contamination in critically ill patients. - Venous arterial blood manipulation system and atrial line contamination.

Emergency and Critical Care Medicine Tokushima University Hospital0 个研究点目标入组 100 人开始时间: 2009年8月6日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
发起方
入组人数
100

研究概览

简要总结

We analyzed microbial contamination for a total of 216 arterial catheters: 109 in the test group and 107 in the comparator group. We found no difference in the incidence of catheter-tip colonization in the two groups (test group 8/109 vs. comparator group 11/107, P = 0.48). There was a statistically significant correlation between catheter-tip colonization and duration of arterial line utilization. We found statistically significantly lower intraluminal fluid contamination in the test group (test group 2/109 vs. comparator group 9/107, P = 0.03). There was no relationship between intraluminal fluid contamination and catheter-tip contamination.

研究设计

研究类型
Interventional

入排标准

年龄范围
18years-old 至 ot applicable(—)
性别
All

入选标准

  • 未提供

排除标准

  • Accidental removal of the catheter Use of <24 h or no aspiration of blood through the arterial line.