SIRONA Trial Heart Failure NYHA Class III
- Conditions
- Heart Failure NYHA Class III
- Interventions
- Device: Cordella™ Heart Failure System
- Registration Number
- NCT03375710
- Lead Sponsor
- Endotronix, Inc.
- Brief Summary
This is a prospective, multi-center, open-label, single-arm feasibility trial to assess device safety and efficacy of the Cordella™ Heart Failure System in 10 NYHA Class III heart failure patients who will receive the Cordella™ Sensor implant.
- Detailed Description
The objectives of this study are to establish that the Cordella™ Sensor can be safely delivered, deployed, and remain stable within the target pulmonary artery (PA) segment through 30 days post-implant as well as to measure the accuracy of Cordella™ Sensor PA pressure measurements compared with fluid-filled catheter PA pressure measurements obtained by standard right heart catheterization (RHC) at 90 days post implant. Safety measures will include the frequency and rates of adverse events, both overall and for each specific event, which will be collected throughout the study. Subjects will remain in this trial for 24 months.
Recruitment & Eligibility
- Status
- ACTIVE_NOT_RECRUITING
- Sex
- All
- Target Recruitment
- 15
- Subject has given written informed consent
- Male or female, at least 18 years of age
- Diagnosis of HF ≥ 3 months with either preserved or reduced left ventricular ejection fraction (LVEF)
- Diagnosis of NYHA Class III HF at the time of Screening with at least 1 HF-related hospitalization, HF treatment in a hospital day-care setting, or urgent outpatient clinic HF visit within the last year
- Subjects must be receiving appropriate medical management of HF for 3 months prior to Screening and clinically stable for at least 1 month prior to study entry
- Subjects with a Body Mass Index (BMI) < 35 kg/mL2 and chest circumference < 1050 mm
- Subjects with right pulmonary artery branch diameter sized between 14 mm and 20 mm (target Sensor implant site) over a length of at least 30mm
- Subjects with distance between target implant site (right PA branch) and ventral thoracic skin surface of < 12cm
- Subjects who are physically able to hold the myCordella™ Patient Reader unit (approximate weight 600g) against the ventral thoracic surface for up to 2 minutes per day while in a seated. position, as well as dock and undock the myCordella™ Patient Reader
- Subjects with sufficient eyesight, hearing, and mental capacity to respond to the myCordella™ Patient Reader's audio/visual cues and operate the myCordella™ Patient Reader.
- Subject has sufficient Cellular and/ or Wi- Fi Internet coverage at home
- Subject agrees to return to the treating Investigator for all scheduled follow up visits and can return to the hospital for follow up
- Subject compliance as defined by data collection and transmissions at least 5 days a week (does not have to be consecutive days) using the myCordella™ Hub & Peripherals for at least 1 week during the screening period
- Subjects with primary pulmonary hypertension.
- Subjects with an active infection at the Sensor Implant Visit
- Subjects with history of recurrent (> 1) pulmonary embolism or deep vein thrombosis
- Subjects who have had a major cardiovascular event (e.g., myocardial infarction, stroke) within 2 months of the Screening Visit
- Subjects whereby RHC or computed tomography pulmonary angiography (CTPA) is contraindicated
- Subjects with a Cardiac Resynchronization Device (CRT), Internal Cardiac Defibrillator (ICD) or pacemaker (if implanted less than 6 months or has more than a single lead), or Left Ventricular Assist Device (LVAD)
- Any major surgery within 30 days of the Sensor Implant Visit.
- Subjects with a Glomerular Filtration Rate (GFR) <30 ml/min or who are on chronic renal dialysis
- Specific liver enzymes [Aspartate Aminotransferase (AST) (SGOT), and Alanine Aminotransferase (ALT) (SGPT) >3 times the upper limit of normal
- Subjects likely to undergo lung and/ or heart transplantation within 24 months of the Screening Visit.
- Subjects with congenital heart disease or mechanical/tissue right heart valve(s)
- Subjects with known coagulation disorders
- Subjects with a hypersensitivity or allergy to platelet aggregation inhibitors including aspirin, clopidogrel, prasugrel, and ticagrelor; or patients unable to take dual antiplatelet or anticoagulants for one month post implant
- Subjects enrolled in another investigational trial.
- Known history of life threatening allergy to contrast dye.
- Subjects who are unwilling or deemed by the Investigator to be unwilling to comply with the study protocol, or subjects with a history of non-compliance
- Subjects with large metal implants/prostheses or any other materials in the thoracic cavity that may interfere with the Cordella™ Pulmonary Artery Sensor System performance
- Subjects with an indwelling inferior vena cava (IVC) filter
- Subjects who are pregnant or breastfeeding
- Subjects whose clinical condition, in the opinion of the Investigator, makes them an unsuitable candidate for the study
- Severe illness, other than heart disease, which would limit survival to <1 year
- Subjects who have had intractable arrhythmias within 2 month of the Screening Visit
- Subjects with patent foramen ovale (PFO)
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Cordella™ Heart Failure System Cordella™ Heart Failure System Cordella™ Heart Failure System and implant of Cordella™ Pulmonary Artery Sensor System (CorPASS)
- Primary Outcome Measures
Name Time Method Efficacy: Accuracy 90 days Accuracy of Cordella™ Sensor PA pressure measurements, compared to standard-of-care fluid-filled catheter PA pressure measurements obtained by standard Right Heart Catheter
Safety Freedom form Adverse events 30 days Freedom from adverse events associated with use of the Cordella™ HF System
- Secondary Outcome Measures
Name Time Method Pressure sensor failure rate 30 days and 2 years Pressure sensor failure rate throughout the study
Device/system-related complications 30 days and 2 years Device/system-related complications
Accuracy of Cordella™ sensor pressure measurements 2 years Accuracy of Cordella™ Sensor PA pressure measurements compared with echo pulmonary artery pressure (PAP) measurements
Frequency of Adverse Events 30 days and 2 years Frequency of adverse events throughout the study
Percentage of device success 90 days Percentage of device success as documented by ability of the System to successfully transmit collected data to a secure database
Heart Failure (HF) Hospitalizations 90 days Frequency of HF hospitalizations, HF treatments in a hospital day-care setting, or urgent outpatient clinic HF visits.
Quality of Life 90 days and 2 years Quality of Life measured by EuroQOL EQ-5D-5L
Change in Pulmonary Artery (PA) pressure 90 days Change in PA pressure pre- and post-implant
Adherence to regular myCordella™ Peripherals measurements 90 days and 2 years Adherence to regular myCordella™ Peripherals measurements
Trial Locations
- Locations (2)
Ziekenhuis Oost Limburg
🇧🇪Genk, Belgium
Galway University Hospital
🇮🇪Galway, Ireland