MedPath

Predictability of Distalization and Derotation of the Carriere Motion Appliance.

Not Applicable
Completed
Conditions
Dental Malocclusion
Interventions
Device: Carriere Motion Appliance®
Registration Number
NCT05094973
Lead Sponsor
University of Salamanca
Brief Summary

To analyze the distalization or distal tooth displacement and derotation angle produced by Carriere Motion Appliance® (CMA) by using a repeatable and reproducible measurement digital technique.

Detailed Description

15 patients with Class II molar and canine relationship were submitted to orthodontic treatment with CMA. All patients were exposed preoperatively (STL1) and postoperatively (STL2) to a digital impression and afterwards were uploaded into a specific cephalometric software to allow the automatic mesh networks alignment of the STL digital files. Subsequently, the distal tooth displacement of upper canines and first upper molars and also the derotation angle of first upper molars were measured. The repeatability and reproducibility were analyzed by Gage R\&R statistical analysis.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
15
Inclusion Criteria
  • Growing patients
  • Late mixed dentition or permanent dentition
  • Fully erupted definitive molars
  • No history of previous orthodontic treatment
  • Complete pre- and post-treatment records (digital study models)
  • Dental Class II relationship.
Exclusion Criteria
  • Permanent dentition
  • Adults
  • Dental Class I, III

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Canine and molar distalization and molar derotationCarriere Motion Appliance®Evaluating the canine and molar distalization and molar derotation in dental Class II patients who where submitted to Carriere Motion Appliance® for a mean of 4 months.
Primary Outcome Measures
NameTimeMethod
Mean Canine Distalization4 months

1.193 mm

Mean Molar Derotation4 months

6.333º

Mean Molar Distalization4 months

1.34 mm

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

University of Salamanca

🇪🇸

Salamanca, Castilla Leon, Spain