临床试验/NCT01917851
NCT01917851Unknown不适用

Evaluate the Accuracy of the NBM200 for Noninvasive Hemoglobin Measurements on Obstetric/Gynecological Patients.

OrSense, Ltd.0 个研究点目标入组 300 人开始时间: 2011年4月最近更新:
适应症

试验速览

阶段
不适用
发起方
OrSense, Ltd.
入组人数
300
主要终点
Compare the Hb values from the NBM200 series against the Hb from the lab blood analyser on obstetric/gynecological patients.

研究概览

简要总结

The purpose of the study is to define the precision, accuracy and variability of the NBM-200 system in non invasive blood Hemoglobin (Hb) measurements relative to the laboratory blood analyzer by comparing two consecutive Hb values obtained by the NBM system to venous values.

详细描述

Each subject will be placed in a comfortable position in a temperature regulated room. Participants will have two consecutive non invasive (NBM) measurements taken which involves inserting the thumb into a ring shaped sensor. After each non invasive measurement the sensor will be removed. Each measurement takes approximately 85 seconds and involves the finger cuff inflating (to over systolic pressure) and deflating several times in quick successions in order to measure and calculate the Hb values. The two non invasive measurements will be performed using the same finger. After the NBM system testing is complete and recorded, the subject will have a venous (venipuncture) blood sample collected in order to provide two blood tests performed on two different models of routine automated CBC or Hb blood analyzers.

研究设计

研究类型
Observational
时间视角
Prospective

入排标准

年龄范围
18 Years 至 55 Years(Adult)
性别
Female
接受健康志愿者

入选标准

  • Female volunteers aged 18 years and over.
  • Any patient subject to CBC testing according to OBGYN standard care and clinical judgment.

排除标准

  • Participants under the age of 18 years
  • Significant deformity, degenerative changes or edema of the thumb or index fingers
  • Localized infection, ulceration or skin breaks involving the fingers
  • Low peripheral body temperature (tissue perfusion) <36.0˚C
  • Vascular disease or Raynaud's phenomenon affecting the fingers
  • Participants who are unable to give informed written consent

结局指标

主要结局

Compare the Hb values from the NBM200 series against the Hb from the lab blood analyser on obstetric/gynecological patients.

时间窗: 1 day

次要结局

未报告次要终点

研究者

发起方
OrSense, Ltd.
申办方类型
Industry
责任方
Sponsor