Validation of the Masimo INVSENSOR00063 in Febrile and Afebrile Subejcts

Masimo Corporation2 个研究点 分布在 1 个国家目标入组 60 人开始时间: 2023年2月23日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
60
试验地点
2
主要终点
Accuracy of Temperature Measurements Using Masimo INVSENSOR00063

研究概览

简要总结

This is a prospective, non-randomized data collection study to evaluate the performance of the Masimo INVSENSOR00063 in obtaining temperature measurements.

详细描述

Individual sites are only representative of a subset of the overall population, per ISO 80601-2-56:2017(Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement), and therefore clinical bias and limits of agreement should not be evaluated for the subsets. Publication of results should include repeatability from individual sites and the combined clinical bias and limits of agreement (NCT05787782, NCT05674344 and NCT05779397).

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
1 Year 至 —(Child, Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Participant age ≥ 1 year
  • Afebrile or febrile at time of enrollment: For febrile subjects, ≥ 99.5 F/ 37.5 C for sublingual temperature reference; ≥ 100.4 F/ 38 C for temporal artery temperature reference.
  • English-or Spanish-speaking patient or parent/Legal Guardian

排除标准

  • Participants deemed not suitable for the study at the discretion of the investigator

结局指标

主要结局

Accuracy of Temperature Measurements Using Masimo INVSENSOR00063

时间窗: About 15-30 minutes

Masimo INVSENSOR00063 temperature measurements will be compared to reference temperature measurements. Accuracy of Masimo INVSENSOR00063 will be measured using clinical bias, limits of agreement and repeatability.

次要结局

未报告次要终点