Precise Coil Positioning in Navigated Transcranial Magnetic Stimulation (nTMS) in Medication- Resistant Major Depressive Disorder (MDD): A Feasibility Study
试验速览
- 阶段
- 不适用
- 状态
- 已完成
- 发起方
- 入组人数
- 30
- 试验地点
- 3
- 主要终点
- Percentage of successful nTMS treatment sessions
研究概览
简要总结
The objective of this study is to evaluate the feasibility of adding a navigational system to traditional repetitive Transcranial Magnetic Stimulation (rTMS, referred to in this application as nTMS) as a way to establish and maintain precise coil positioning (contact, rotation, and tilt) and consistent brain region targeting throughout a nTMS treatment session and in subsequent nTMS sessions.
详细描述
Objective
The objective of this study is to evaluate the feasibility of adding a navigational system to traditional repetitive Transcranial Magnetic Stimulation (rTMS, referred to in this application as nTMS) as a way to establish and maintain precise coil positioning (contact, rotation, and tilt) and consistent brain region targeting throughout a nTMS treatment session and in subsequent nTMS sessions. Success will be evaluated via data analysis of information captured during nTMS, which will evaluate coordinates, contact, rotation, and tilt parameters of the stimulation pulses delivered. The study will be conducted with patients who have been selected as traditional rTMS candidates and have met the criteria for rTMS, which includes, but is not limited to the following: Adults with Major Depressive Disorder (MDD) who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode. For the purposes of this study, nTMS is referring to the use of a navigation device in combination with delivery of traditional rTMS.
Aim 1: Determine the percentage of successful nTMS treatment sessions. A successful nTMS treatment session is defined as the following: 80% of the TMS pulse trains are delivered while the coil is within acceptance criteria measured by tracking the following parameters: coil deviations from digitally saved target location, which includes contact (mm), rotation (°), and tilt (°) parameters. The acceptance criteria are met with coil position at the time of pulse delivery as follows: coil rotation within ±15° of target; coil tilt within ±15° of target; in-plane deviation within ±5mm of target; contact deviation within ±2mm of target.
Aim 2: Determine Patient Health Questionnaire (PHQ-9) change measured weekly from baseline to after 30 sessions of nTMS, collect Patient comfort data, measure Operator confidence and collect coil position data for nTMS sessions to determine future product improvements and clinical studies.
The goal of this proposal is to establish nTMS as a means of precise administration of transcranial magnetic stimulation. Ultimately, data from this study will be used to design larger, comparative studies to establish superior efficacy profiles in the treatment of MDD.
研究设计
- 研究类型
- Interventional
- 分配方式
- Na
- 干预模型
- Single Group
- 主要目的
- Treatment
- 盲法
- None
入排标准
- 年龄范围
- 18 Years 至 80 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Diagnosis of medication-resistant MDD,
- •Age 18 years or older
- •Normal findings in the medical history, physical, and neurological examination
排除标准
- •History of seizure disorder
- •History of neuroleptic medications/prior use of neuroleptics
- •Presence of implanted medical pump, metal plate, or metal object in skull or eye
- •Pregnant women
结局指标
主要结局
Percentage of successful nTMS treatment sessions
时间窗: The number of successful treatment sessions will be assessed after 6 weeks of treatment. Only data from week 2 through 6 will be included (approximately 25 treatment sessions). (Example 20 out of 25 sessions were successful.)
The primary outcome measure will be the percentage of successful nTMS treatment sessions pooled from sessions 6 through 30 for each study subject. A successful nTMS treatment session is defined as: 80% of the TMS pulse trains are delivered while the coil is within acceptance criteria measured by tracking the following parameters: coil deviations from digitally saved target location, which includes contact (mm), rotation (°), and tilt (°) parameters. The acceptance criteria are met with coil position at the time of pulse delivery as follows: coil rotation within ±15° of target; coil tilt within ±15° of target; in-plane deviation within ±5mm of target; contact deviation within ±2mm of target.
次要结局
- Patient Health Questionnaire 9 (PHQ-9) - Patient Depression Questionnaire(At baseline and after six weeks of treatment)
- Patient Comfort(After last treatment has been completed; approximately six week after treatment start)
- Operator Confidence(Aproximately 5 months after first patient treatment. After last study patient at site has completed the treatment or has exited the study (expected for May 2019).)
- Coil Position deviation from target(Coil position deviation: every treatment session during week 2 through 6)