Dynamic FDG PET/CT: Optimization and Validation of Data Acquisition

Not Applicable
Completed
Conditions
Positron-Emission Tomography
Interventions
Diagnostic Test: Dynamic Whole-Body parametric PET/CT
Registration Number
NCT04660279
Lead Sponsor
Aarhus University Hospital
Brief Summary

Quantification of the metabolic rate of glucose from Dynamic Whole-Body PET examinations requires measurements of the time course of the radioactivity concentrations in arterial blood by blood sampling, and in the tissue of interest by dynamic PET. Invasive arterial blood sampling cannot be part of a standard examination, and therefore the blood samples need to be replaced by activity concentrations derived from the PET images, usually from small volumes in the descending aorta or left ventricle.

Newly developed scanner software (Siemens) allows automated CT-based identification of blood pool regions and extraction of an image-derived blood input function from the corresponding PET data.

However, this automated method needs validation, as it could be prone to systematic errors caused by limited spatial resolution, patient movement, and image reconstruction. We will use invasively measured arterial blood samples as a reference for validation of methods to extract non-invasive PET image-derived input functions and quantify any systematic errors that could propagate to the resulting parametric images.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
20
Inclusion Criteria
  • Patients with a planned clinical standard PET/CT as part of pre-therapeutic staging, response assessment scanning.
  • Good performance status, which will allow us to obtain informed consent to draw the required arterial blood samples during the examination, and which will permit patients to lay still in the scanner for at least 70 min.
Exclusion Criteria
  • • Age < 18 or > 85 years

    • Patients that can not tolerate a PET scan (f.ex: claustrophobia).

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
ValidateDynamic Whole-Body parametric PET/CTD-WB PET/CT scans + arterial blood sampling.
Primary Outcome Measures
NameTimeMethod
Time reduction of current parametric reconstruction protocolthrough study completion, an average of 1 year

After we have validated the image derived input function we will use the data we sampled to construct a "normal population" input function that can be applied to the current protocol. This way we will be able to skip the acquisition on the camera of the input function and be able to only scan the patients at a later period, therefor shortening the acquisition protocol to something more manageable in daily clinic.

Validation of current parametric reconstruction protocolthrough study completion, an average of 1 year

We will perform a correlation of the image derived input function and the input function obtained from blood sampling. This will allow us to confirm that the information obtained by the equipment is accurate enough and can be trusted.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

Aarhus University Hospital

🇩🇰

Aarhus, Denmark