Verification Protocol for The Cloud DX Vitaliti™ Continuous Vital Sign Monitor (CVSM): Non-invasive Blood Pressure Component for Clinical Use and Self Measurement According to the American Safety in Medical Technology and the IEEE

Population Health Research Institute2 个研究点 分布在 1 个国家目标入组 78 人开始时间: 2018年8月9日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
78
试验地点
2
主要终点
Blood pressure

研究概览

简要总结

The purpose of this study is the verification of Cloud-DX Vitaliti™ Continuous Automated Non-Invasive Blood Pressure Monitor for Clinical Use and Self Measurement according to standards laid out by IEEE Std 1708 and AAMI-ISO 81060-2.

详细描述

Several automated vital sign measurement systems have been developed recently; however, non-invasive continuous blood pressure monitoring continues to be a major limitation. To circumvent the limitations of current systems of interval hemodynamic measurement and increasing demands on health care providers, a non-invasive automated continuous vital sign monitor that can integrate into hospital early warning systems and warn health care providers of deteriorating vital sign parameters would be of significant value and a major advance for current hospital monitoring systems.

Cloud-DX Vitaliti™ is a new low profile automated vital sign monitor, which includes a continuous non-invasive blood pressure measurement system. Additional parameters are also measured. Paired with companion software from cloud based computing, Vitaliti™ features a user friendly interface, onboard radios for communication with smartphones/tablets/personal computers, and several sensors.

Cloud-DX Vitaliti™ has undergone preliminary testing and is now ready for verification to achieve compliance with AAMI/ISO and IEEE standards of medical grade device. At the Hamilton General Hospital (HGH), Cardiac Surgical Intensive Care Unit (CICU) and Vascular Surgical Step Down Units, arterial catheters are standard of care and performed on most patients for variable periods of post-operative recovery. Within this setting, the IEEE Std 1708 and AAMI-ISO 81060-2 would be fully met. Furthermore, the cardiovascular surgical caseloads at HGH would provide sufficient test subjects to validate VitalitiTM in a timely fashion.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Patients ≥18 years of age;
  • Patient has undergone major cardiac or vascular surgery and has been admitted post-op to the cardiac surgical intensive care unit (CICU) or vascular surgery step down unit with an arterial line in place

排除标准

  • Patients < 18 years of age;
  • Current arrhythmia
  • Blood pressure outside of test range for phase (refer to table 1)
  • Allergy to device contact adhesive (e.g. ECG leads)

结局指标

主要结局

Blood pressure

时间窗: 30 minutes

Determining the accuracy of non-invasive blood pressure measurement compared to invasive blood pressure. A detailed analysis will be conducted on the invasive blood pressure to verify accuracy as defined in the ISO 81060-2 Standard.

次要结局