跳至主要内容
临床试验/NCT03843242
NCT03843242已完成不适用

Evaluation of Ventricular Pacing Suppression Algorithms in Dual Chamber

Keimyung University Dongsan Medical Center6 个研究点 分布在 1 个国家目标入组 146 人开始时间: 2019年5月30日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
146
试验地点
6
主要终点
Percentage of right ventricular pacing

研究概览

简要总结

The recent study using IRSplus and VpS algorithm from Biotronik pacemaker showed the significant reduction in ventricular pacing to less than 3%.

The purpose of this study is to evaluate the efficacy of IRSplus and VpS algorithm in reducing ventricular pacing compared with conventional DDD pacing with a fixed AV delay.

详细描述

The adverse cardiac outcomes due to right ventricular apical pacing with dual chamber pacemakers have been widely observed such as ventricular dyssynchrony resulting in reduced left ventricular function, increased risk of heart failure and atrial fibrillation. To minimize the ventricular pacing, manufacturers of pacemaker have made an effort to develop special algorithms designed to deliver right ventricular pacing only in case of demonstrated persistent long PR interval or repetitively lacking intrinsic ventricular activation.

Currently, there are three methods and algorithms are available with a pacemaker from the Biotronik SE & Co. KG to minimize right ventricular pacing as follows; DDD mode with fixed longer atrioventricular (AV) delay than intrinsic conduction time; Intrinsic rhythm support (IRSplus); Ventricular pacing suppression (VpS).

In the real world, it is the most common practice to program the DDD(R) mode with fixed long AV delay because of physician's concerns about the possible failure of an algorithm or long pause being resulted from 2 consecutive loss of AV conduction by the algorithm. However, the main disadvantage of a fixed with long AV delay is that the prolonged total atrial refractory period (TARP) results in changes of the upper rate behavior (i.e., pseudo-Wenckebach AV block and subsequently 2:1 block at lower atrial tracking rates). The possibility of pacemaker-mediated tachycardia is getting high if the post-ventricular atrial refractory period (PVARP) is shortened to compensate. Furthermore, there is a high chance to have fusion/pseudo-fusion of ventricular pacing in a fixed long AV delay because the AV conduction is dynamic according to the heart rate.

The recent systematic review showed that there are no significant differences between the pacing modes for mortality, heart failure, stroke, and atrial fibrillation (AF) in patients with sinus node dysfunction (SND) without AV block. However, the dual chamber pacemaker is still recommended in patients with SND due to lack of tools to identify patients at high risk of developing the complete AV block. The meta-analysis about the effect of the reduction in unnecessary ventricular pacing using a sophisticated algorithm in patients with SND showed there are no benefits in clinical outcomes compared with conventional DDD mode. But the percentage of ventricular pacing in ventricular pacing reduction modality group was not negligible between 1~11.5%. The result might be changed if we have data with a more significant reduction in ventricular pacing using difference algorithm. The recent study using IRSplus and VpS algorithm from Biotronik pacemaker showed the significant reduction in ventricular pacing to less than 3%.

The purpose of this study is to evaluate the efficacy of IRSplus and VpS algorithm compared with conventional DDD pacing with a fixed AV delay.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Diagnostic
盲法
Single (Participant)

入排标准

年龄范围
20 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
否

入选标准

  • •Sick sinus syndrome
  • •No evidence of 2nd and 3rd degree AV block
  • •Provide written informed consent
  • •Age ≥ 20 years old

排除标准

  • •2nd and 3rd degree AV block
  • •History of AF
  • •patients with older version of pacemaker
  • •Life expectancy ≤ one year
  • •Pregnant or lactating women

研究组 & 干预措施

Pacemaker with Fixed long AV delay

Experimental
  1. Patients who meet the inclusion criteria and is implanted with a Biotronik Enitra 8 DR-T pacemaker are eligible.
  2. The pacemaker was programmed with a long and fixed atrioventricular interval for the first 3 months.
  3. Definition of fixed AV delay (than intrinsic AV conduction) • If P-wave exists: intrinsic AV conduction time = As ~ Vs interval in the marker channel sensed AV delay = intrinsic AV conduction time + 20 msec paced AV delay = sensed AV delay + 30 msec • If no P-wave exits: intrinsic AV conduction time = Ap ~ Vs interval in the marker channel paced AV delay = intrinsic AV conduction time + 20 msec sensed AV delay = paced AV delay - 30 msec • If the intrinsic AV conduction time is ≥ 300ms, make paced/sensed AV delay 350/320 msec

干预措施: Pacemaker with Fixed long AV delay (Device)

Pacemaker with VpS® algorithm on

Experimental

Vp Suppression ON algorithm: This feature promotes the intrinsic AV conduction by only pacing the ventricle when intrinsic conduction becomes unstable or disappears. Depending on the presence or absence of AV conduction, the feature is implemented either in the ventricular pacing suppression state ADI(R), which promotes the intrinsic conduction, or in the DDD(R) ventricular pacing state Vp DDD(R), which provides ventricular pacing. Automatic switching capabilities between those two states promotethe intrinsic conduction as much as possible without harming the patient. Scheduled Vs searching tests look for intrinsic conduction using an extended AV delay of 450ms.

干预措施: Pacemaker with Fixed long AV delay (Device)

Pacemaker with VpS® algorithm on

Experimental

Vp Suppression ON algorithm: This feature promotes the intrinsic AV conduction by only pacing the ventricle when intrinsic conduction becomes unstable or disappears. Depending on the presence or absence of AV conduction, the feature is implemented either in the ventricular pacing suppression state ADI(R), which promotes the intrinsic conduction, or in the DDD(R) ventricular pacing state Vp DDD(R), which provides ventricular pacing. Automatic switching capabilities between those two states promotethe intrinsic conduction as much as possible without harming the patient. Scheduled Vs searching tests look for intrinsic conduction using an extended AV delay of 450ms.

干预措施: Pacemaker with VpS® algorithm on (Device)

Pacemaker with IRSplus algorithm on

Experimental

IRS plus algorithm: This algorithm incorporates two different functions: the first is scan hysteresis, which better enables the heart to pace on its own by periodically extending the search time for its natural pacing stimulus (the intrinsic AV conduction) over six consecutive atrial cycles. The second is the repetitive hysteresis, which recognizes when the heart is not pacing on its own (a consistent loss of intrinsic AV conduction lasting for six consecutive atrial cycles) and switches the mode of the device from extended to basic atrioventricular (AV) delay.

干预措施: Pacemaker with Fixed long AV delay (Device)

Pacemaker with IRSplus algorithm on

Experimental

IRS plus algorithm: This algorithm incorporates two different functions: the first is scan hysteresis, which better enables the heart to pace on its own by periodically extending the search time for its natural pacing stimulus (the intrinsic AV conduction) over six consecutive atrial cycles. The second is the repetitive hysteresis, which recognizes when the heart is not pacing on its own (a consistent loss of intrinsic AV conduction lasting for six consecutive atrial cycles) and switches the mode of the device from extended to basic atrioventricular (AV) delay.

干预措施: Pacemaker with IRSplus algorithm on (Device)

结局指标

主要结局

Percentage of right ventricular pacing

时间窗: During 10 ~ 15 months after the study enrollment

the right ventricle pacing percentage appeared on the pacemaker interrogation

Percentage of right ventricular pacing

时间窗: During first 3 months after the study enrollment

the right ventricle pacing percentage appeared on the pacemaker interrogation

Percentage of right ventricular pacing

时间窗: During 4 ~ 9 months after the study enrollment

the right ventricle pacing percentage appeared on the pacemaker interrogation

次要结局

  • Percentage of atrial high rate episode(During 12 months after randomization (15 months after enrollment))
  • New onset atrial fibrillation(During 12 months after randomization (15 months after enrollment))
  • Occurrence of heart failure(During 12 months after randomization (15 months after enrollment))

研究者

申办方类型
Other
责任方
Principal Investigator
主要研究者

Seongwook Han

Professor of Internal Medicine

Keimyung University Dongsan Medical Center

研究点 (6)

Loading locations...

相似试验