OM2 Low Saturation Verification
- Conditions
- Hypoxia
- Registration Number
- NCT06661889
- Lead Sponsor
- Medtronic - MITG
- Brief Summary
The verification of an investigational pulse oximetry board to verify pulse rate and saturation accuracy over a specified saturation range in diverse populations.
- Detailed Description
To verify the SpO2 and pulse rate accuracy in comparison to a reference standard blood measurements. To achieve paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of the prototype pulse oximeter on a group of healthy adult volunteers. The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steadystate saturation periods. In this study, the investigational pulse oximetry PCBA system saturation measurements will be compared to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from a diverse pool of healthy human subjects.
Recruitment & Eligibility
- Status
- RECRUITING
- Sex
- All
- Target Recruitment
- 24
- Healthy volunteers 18 to ≤39 years
- Subject is willing and able to comply with study procedures and duration
- Subject is willing to sign an informed consent
- Subject weighs >40kg
- Subject is a non-smoker or has not smoked within 2 days prior to the study
- Cleared same day health assessment form and health screening
-
Subject is considered as being morbidly obese (defined as BMI >39.5)
-
Compromised circulation, injury, or physical malformation of fingers, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
-
Subjects of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study
-
Subjects with known concurrent, active respiratory conditions such as:
- uncontrolled / severe asthma
- flu or influenza type infection
- pneumonia / bronchitis
- shortness of breath / respiratory distress
- unresolved respiratory or lung surgery
- emphysema, COPD, lung disease
- recent COVID (last 2 months)
-
Subjects with known heart or cardiovascular conditions such as:
- hypertension: systolic >140mmHg, or Diastolic >90mmHg on 3 consecutive readings
- have had cardiovascular surgery
- chest pain (angina)
- heart rhythms other than a normal sinus rhythm and/or with respiratory sinus arrhythmia, at the discretion of the Principal Investigator or designee
- previous heart attack
- blocked artery
- unexplained shortness of breath
- congestive heart failure (CHF)
- history of stroke
- transient ischemic attack
- carotid artery disease
- myocardial ischemia
- myocardial infarction
- cardiomyopathy
- implantable active medical device such as pacemaker or automatic defibrillator
-
Self-reported health conditions as identified in the Health Assessment Form
- diabetes
- uncontrolled thyroid disease
- kidney disease / chronic renal impairment
- history of seizures (except childhood febrile seizures)
- epilepsy
- history of unexplained syncope
- recent history of frequent migraine headaches
- recent symptomatic head injury, within the last 2 months
- cancer requiring chemotherapy, radiation, or current treatment
- subjects with known clotting disorders
- history of bleeding disorders or personal history of prolonged bleeding from injury
- history of blood clots
- hemophilia
- sickle cell trait or disease
- current use of blood thinner: prescription or daily use such as aspirin
- subjects with Severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors
- participants with severe allergy to iodine (only applicable if iodine is used)
- subjects with prior or known severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) or heparin
- arterial cannulation within the last 30 days prior to study date, (this may exclude only one radial artery site, left or right)
- history of clinically significant complications from previous arterial cannulation
- unwillingness or inability to remove colored nail polish or colored artificial nails other than clear from test digits
- other known health condition, upon disclosure in Health Assessment form
Study & Design
- Study Type
- OBSERVATIONAL
- Study Design
- Not specified
- Primary Outcome Measures
Name Time Method Acceptance criteria for pulse rate accuracy within 20-250 BPM 2.5 hours Pulse rate accuracy Max-A \<=3 BPM (ARMS) Max-N \<=3 BPM (ARMS) OxySoft-N \<=3 BPM (ARMS) Max-Fast \<=3 BPM(ARMS)
Verification of Saturation Accuracy 2.5 hours Verification of Saturation Accuracy Saturation accuracy in a diverse subject population over a specified saturation range.
The acceptance criteria for SpO2 during low saturation conditions across the saturation range of 60-80%:
SpO2 Max-A ±3% (ARMS) Max-N ±3% (ARMS) OxySoft-N ±3% (ARMS) Max-Fast ±3%(ARMS)Acceptance criteria for SpO2 accuracy across the saturation range of 70-100%: 2.5 hours Verification of Saturation Accuracy Saturation accuracy in a diverse subject population over a specified saturation range. The acceptance criteria for SpO2 during saturation conditions across the saturation range of 70-100%: SpO2 Max-A ±2% (ARMS) Max-N ±2% (ARMS) OxySoft-N ±2% (ARMS) Max-Fast ±2%(ARMS)
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Medtronic Clinical Physiology Lab
🇺🇸Denver, Colorado, United States