Elimination of CO2-insufflation-induced Hypercarbia in Open Heart Surgery
- Conditions
- Adult Patients Undergoing Open Heart Replacement of the Aortic Valve for Aortic Stenosis or Insufficiency
- Interventions
- Procedure: Tube clamping
- Registration Number
- NCT04202575
- Lead Sponsor
- Karolinska University Hospital
- Brief Summary
The study aim was to evaluate if an additional separate venous reservoir eliminates CO2-insufflation induced hypercapnia and keeps sweep gas flow of the oxygenator constant during open heart surgery.
- Detailed Description
Background: CO2-gas insufflation is used for continuous de-airing during open heart surgery. The study aim was to evaluate if an additional separate venous reservoir eliminates CO2-insufflation induced hypercapnea and keeps sweep gas flow of the oxygenator constant.
Methods: A separate small reservoir are used during CPB in addition to a standard large venous reservoir. The small reservoir receive drained wound blood and CO2-gas continuously via a suction drain (1 L/min) and handheld suction devices from the open surgical wound. CO2-gas is insufflated via a gas-diffuser in the open surgical wound at 10 L/min. During cross-clamping, gas and blood are either continuously drained to the standard large venous reservoir or not, every 5 minutes after steady state of PaCO2 is observed, after adjustment of sweep gas flow as necessary. Mean values for each setup (2-4 times) for each patient will be analyzed with Wilcoxon rank-sum test.
Recruitment & Eligibility
- Status
- UNKNOWN
- Sex
- All
- Target Recruitment
- 10
- adult patients undergoing open heart replacement of the aortic valve
- Use of perioperative cardiopulmonary bypass
- Denied participation in the study
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- CROSSOVER
- Arm && Interventions
Group Intervention Description Intervention setup Tube clamping The connecting tube between the additional and standard venous reservoir is clamped. Thus, blood and gas from the coronary and cardiotomy suction devices are collected in the additional venous reservoir. During the intervention setup, the blood in the additional venous reservoir is only evacuated to the standard reservoir if the volume exceeded 800ml, and always with a remaining volume of 100 mL blood to keep the CO2-gas trapped in the additional venous reservoir.
- Primary Outcome Measures
Name Time Method Difference in PaCO2 (kPa) At 5 minutes steady state
- Secondary Outcome Measures
Name Time Method Oxygenator sweep gas flow rate (L/min) At 5 minutes steady state
Trial Locations
- Locations (1)
Karolinska University Hospital
🇸🇪Stockholm, Sweden