"Low Flow" CO2 Removal on RRT
- Conditions
- Renal Replacement Therapy
- Interventions
- Device: Extracorporeal CO2 removal with the Prismalung(R) gas exchanger on the Prismaflex(R) renal replacement platform
- Registration Number
- NCT02590575
- Lead Sponsor
- University of Erlangen-Nürnberg Medical School
- Brief Summary
The purpose of this study is to test the effectiveness of a membrane gas exchange device in the venovenous circulation of a continuous renal replacement therapy for the purpose of CO2 elimination and pH compensation. Thus, the primary endpoint is the modification of the PaCO2 and/or the ventilator settings (tidal volume VT and plateau pressure Pplat).
- Detailed Description
Protocol synopsis Study title "Low flow" CO2 removal via a membrane gas exchange device (Prismalung®) on a renal replacement platform (Prismaflex®) in hypercapnic, ventilated patients requiring renal replacement therapy Brief description Prismalung Study Indication Hypercapnic, ventilated, and renal replacement therapy requiring patients Primary study goal Changes in the arterial pCO2 value Secondary study goals Changes in the acid-base balance Changes of tidal volumes/ventilation pressures Changes in hemodynamics Study design Prospective intervention study (pilot study)
Study population Inclusion criteria:
1. Necessity of renal replacement therapy
2. Necessity of ventilation therapy with an expected duration \>24 hrs. in mild to moderate ARDS according to the "Berlin Definition" (300≥PaO2/FiO2\>100mmHg, ventilated acc. to the ARDS network strategy with VT 6 ml/kgKG (KG = Körpergewicht, body weight) and PEEP ≥ 5 cmH2O)Pneumonia, pulmonary vasculitis, COPD, or other pulmonary disorder
3. PaCO2 ≥ 55 mmHg with plateau ventilation pressure \> 25 cmH2O and pH \< 7.30
4. Written consent by patient or legal representative
Number of patients n = 20 Interventions 1. With ongoing continuous hemofiltration and mechanical ventilation additional membrane gas exchange device (Prismalung®) in the renal replacement circulation; measurements of the effects on PaCO2 under varying blood and sweep gas flows 2. Afterwards, reduction of the tidal volume (VT) / plateau pressure until reaching the initial PaCO2value Study drug None Primary endpoint 1. Changes in the PaCO2 / acid-base status (BGA) 2. Changes of the VT / plateau pressure in the ventilation (ventilation settings) Secondary endpoints Changes in the systemic hemodynamics Vasopressor dosage BGA, renal values, lactate PaCO2 before and after gas exchange device Lifetimes of the extracorporeal circulation Complications Study site Intensive care units Medizinische Klinik 4, Uniklinikum Erlangen-Nürnberg Medizinische Klinik 4, Klinikum Nürnberg Süd Klinik für Intensivmedizin Universitätsklinikum Hamburg Eppendorf Klinik für Anästhesiologie und Intensivmedizin Universitätsklinikum Frankfurt a. M. Timeline Start in October 2015 Anticipated study duration: 6 months Financing Baxter Gambro Renal GmbH (Ltd)
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 20
- Necessity of renal replacement therapy
- Necessity of ventilation therapy with an expected duration >24 hrs. in mild to moderate ARDS according to the "Berlin Definition" (300≥PaO2/FiO2>100mmHg, ventilated acc. to the ARDS network strategy with VT 6 ml/kgKG (KG = Körpergewicht, body weight) and PEEP ≥ 5 cmH2O) Pneumonia, pulmonary vasculitis, COPD, or other pulmonary disorder
- PaCO2 ≥ 55 mmHg with plateau ventilation pressure > 25 cmH2O and pH < 7.30
- Written consent by patient or legal representative
- Age < 18 years
- Pregnancy
- BMI > 40 kg/m2
- Decompensated heart failure or acute stroke
- Severe ARDS (PaO2/FiO2 < 100 mmHg)
- Acute cranio-cerebral injury
- Severe liver failure (Child-Pugh score >7)
- Heparin-induced thrombopenia (HIT II)
- Contraindications to the performance of CVVH (continuous venovenous hemofiltration)
- Contraindications to systemic anticoagulation
- Lacking possibility of access via a dialysis catheter
- Surgeries scheduled within 48 hrs of inclusion in the study
- Advanced tumor disorder with life expectancy < 1 month
- Moribund patients, decision to forego therapy
- Lacking consent
- Participation in a different intervention study
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description CO2 removal Extracorporeal CO2 removal with the Prismalung(R) gas exchanger on the Prismaflex(R) renal replacement platform -
- Primary Outcome Measures
Name Time Method Changes of the VT / plateau pressure in the ventilation (ventilation settings) hours to 3 days Changes in the PaCO2, acid-base status (BGA) hours to 3 days
- Secondary Outcome Measures
Name Time Method Standard Base Excess (mmol/l) hours to 3 days Changes in heart rate (bpm) hours to 3 days Changes in norepinephrin dose (mg/h) hours to 3 days Changes in Serum lactate concentration (mmol/l) hours to 3 days Arterial bicarbonate concentration (mmol/l) hours to 3 days Changes in mean arterial pressure (mmHg) hours to 3 days Arterial pO2 and pCO2 (mmHg) hours to 3 days PCO2 before and after gas exchange device (mmHg) hours to 3 days Number and kind of adverse events during study period deemed clinically significant by investigator (as documented by paper questionnaire) hours to 3 days Changes in Serum creatinine (mg/dl) hours to 3 days Lifetime of extracorporeal circuit in hours (Duration from start to end of RRT) hours to 3 days
Trial Locations
- Locations (1)
University of Erlangen-Nuernberg
🇩🇪Nuernberg, Germany