Parametric Testing of Direct Physiological Effects of Transcutaneous Vagus Nerve Stimulation (tVNS) in Healthy Controls

Phase 1
Completed
Conditions
Healthy
Interventions
Device: Digitmer High Voltage Stimulator model DS7A TENS unit
Registration Number
NCT02835885
Lead Sponsor
Medical University of South Carolina
Brief Summary

The purpose of this study is to determine the feasibility and safety of Transcutaneous Vagus Nerve Stimulation (tVNS) in healthy controls. Tested in healthy subjects (HS) the physiological effects of tVNS at different stimulation parameters. The electrical stimulation variables will be pulse width, frequency, and current. This will be tested in a laboratory setting and will measure tVNS effects on heart rate, respiration rate, blood pressure, and perceived pain. This is a feasibility and small safety study.

Hypothesis: tVNS is feasible and safe

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
15
Inclusion Criteria
  • Meets none of the Exclusion Criteria
Read More
Exclusion Criteria
  • History of seizures, depression, or pain conditions
  • Taking any prescription medications (excluding contraceptives and statins)
  • History of any ear trauma
  • History of heart conditions
  • Has a heart pacemaker or artificial valve
  • Has a metal implant above the waste
  • Is pregnant
  • Alcohol Dependence
  • Illicit drug use in past 6 months
  • History of brain surgery or lesions
  • History of loss of consciousness (>15 min)
  • Faint at the sight of blood
  • Hospitalization or surgeries in past 6 months
Read More

Study & Design

Study Type
INTERVENTIONAL
Study Design
CROSSOVER
Arm && Interventions
GroupInterventionDescription
Sham tVNS StimulationDigitmer High Voltage Stimulator model DS7A TENS unitStimulating electrode will be placed on left ear lobe of subject for sham tVNS stimulation condition. Current will be held constant (200% perceptual threshold) while pulse width and frequency are varied. The intervention used to keep current constant is Digitmer High Voltage Stimulator model DS7A TENS unit. Each stimulation period (9 randomized stimulation parameters varying in pulse width and frequency) will last 1 minute with a 3 minute break between each parameter. Physiological data (O2 saturation, blood pressure, breathing rate, and Electrocardiogram (EKG) to measure heart rate ) will be collected continuously.
Active tVNS StimulationDigitmer High Voltage Stimulator model DS7A TENS unitStimulating electrode will be placed on left tragus of subject. Current will be held constant (200% perceptual threshold) while pulse width and frequency are varied.The intervention used to keep current constant is Digitmer High Voltage Stimulator model DS7A TENS unit. Each stimulation period (9 randomized stimulation parameters varying in pulse width and frequency) will last 1 minute with a 3 minute break between each parameter. Physiological data (O2 saturation, blood pressure, breathing rate, and Electrocardiogram (EKG) to measure heart rate ) will be collected continuously.
Primary Outcome Measures
NameTimeMethod
Change from Baseline O2 Saturation Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 2Total time for Visit 2 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data O2 saturation (%oxygen) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline Heart Rate Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 1Total time for Visit 1 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data,Electrocardiogram (EKG) to measure heart rate in beats/sec will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline Heart Rate Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 2Total time for Visit 2 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data,Electrocardiogram (EKG) to measure heart rate in beats/sec will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline O2 Saturation Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 1Total time for Visit 1 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data O2 saturation (% oxygen) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Secondary Outcome Measures
NameTimeMethod
Change from Baseline Blood Pressure Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 1Total time for Visit 1 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data Blood Pressure (mmHg) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline Blood Pressure Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 2Total time for Visit 2 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data Blood Pressure (mmHg) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline Breathing Rate Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 2Total time for Visit 2 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data breathing rate (%CO2) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.

Change from Baseline Breathing Rate Before Each Stimulation Period (1-9) to Post Stimulation Period (1-9) Visit 1Total time for Visit 1 (either active or sham condition) is 50 minutes - 9 Stimulation Periods (each 4 minutes long with 90 seconds between)

There are nine stimulation periods, each with own stimulation parameters that consist of a constant current, pulse width and frequency. Each stimulation period has 1 minute of stimulation (active or sham), followed by 3 minutes of post stimulation baseline recording Each of the stimulation period last 4 minutes with an additional 90 seconds between each Stimulation period. Current will be 200% of the subject's perceptual threshold for each stimulation parameter. The pulse width will be either 100 microseconds (µs), 200 µs, or 500µs. Frequency will be either 1 Hertz (Hz), 10 Hz, or 25 Hz.

Baseline physiological data breathing rate (%CO2) will be collected before each stimulation parameter, during stimulation parameter, and post each stimulation parameter.