Endurance, Energy Expenditure, Perceived Function, and Satisfaction of Persons With Transtibial Limb Loss Using a Running-Specific Prosthesis Modified for Walking

University of Washington2 个研究点 分布在 1 个国家目标入组 36 人开始时间: 2015年5月最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
36
试验地点
2
主要终点
Six-minute Walk Test (6MWT), Participants With Bilateral Amputation

研究概览

简要总结

The goal of this study is to evaluate endurance, walking performance, mobility, and perceived exertion of transtibial prosthesis users (i.e., study participants) walking with a conventional energy storing prosthetic foot (ESF) and a novel modified running-specific prosthesis (mRSF). A randomized cross-over study will be conducted to determine if the mRSF provides superior performance to the ESF, which is commonly prescribed to most active individuals with lower limb amputation.

详细描述

Amputation of a limb is a life-altering event with profound physical, psychological, and social implications. To address their functional, vocational, and recreational needs, people with lower limb amputation (LLA) are often provided with a prosthesis or artificial leg. While use of a prosthesis can allow an individual to achieve a basic level of functional mobility, absence of an anatomical foot and ankle still impairs their physical performance. As a result, people with LLA regularly exhibit decreased walking speeds, diminished endurance, and restricted ability to participate in desired life situations.

Over the past three decades, increasingly sophisticated prosthetic foot designs have been developed by the prosthetics industry to replace amputated structures in the leg. Contemporary, energy storing feet (ESF) employ advanced materials and unique geometric designs to improve walking performance and endurance of their users. Although prosthetic limbs with ESF allow people with LLA the potential to return to an active lifestyle, even the most advanced ESF do not significantly reduce the increased energy demands required for walking when compared to conventional prosthetic feet.

Commercially-available running-specific feet (RSF) like the Össur Cheetah (Össur, Reykjavik, Iceland) allow people with LLA to participate in athletic activities and sporting events. RSF provide significantly enhanced performance, compared to traditional ESF, by extending the length and increasing the stiffness of the prosthetic keel (forefoot). RSF also do not include a heel, as they are used only for running activities. Although transtibial runners with RSF exhibit endurance levels similar to non-amputees, the RSF design does not allow the biomechanical movements or provide the stability needed to use the foot for walking (over level or uneven terrain).

A novel modified running-specific foot (mRSF) has been developed to integrate the performance of a RSF with the utility of an ESF. The mRSF combines the running keel (forefoot) of a RSF with the walking heel of a ESF. The mRSF can then be used for walking, running, and other routine daily activities. Preliminary feedback on the mRSF suggests that users experience improved overall function and high satisfaction with the device. However, empirical evidence is needed to support prescription of this prosthetic foot at other clinical facilities.

The goal of this study is therefore to evaluate endurance, walking performance, mobility, and perceived exertion of transtibial prosthesis users (i.e., study participants) walking with a conventional ESF and the mRSF. Results will be compared to determine if the mRSF provides superior performance to the ESF, which is commonly prescribed to most active individuals with LLA.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Crossover
主要目的
Treatment
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • unilateral or bilateral, transtibial (below-knee) amputation from non-dysvascular causes
  • Medicare functional classification level (K-level) 3 or higher (unlimited community ambulatory)
  • prosthetic user for > 1 year
  • scheduled to receive a mRSF prosthesis
  • able and willing to participate in the study protocol (e.g., treadmill walking, overground walking, respond to survey questions, able to read and write English)

排除标准

  • contralateral lower or upper limb involvement
  • any health condition that would limit participation in the study procedures (e.g., skin breakdown, heart disease)

结局指标

主要结局

Six-minute Walk Test (6MWT), Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Participants were asked to walk as far as possible in 6 minutes. Participants walked around cones placed 30m apart in a level hallway. Total distance (in meters) walked in 6 minutes was recorded.

Step Length, Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step length (cm) was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT). Step length is reported for both the prosthetic and sound (intact) limbs.

Metabolic Oxygen Consumption (VO2), Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Breath-by-Breath VO2 was collected as participants walked on a treadmill at self selected fast, normal, and slow speeds. Participants walked for 6 minutes at each speed (and in each experimental condition). VO2 was normalized energy expenditure to each participant's biological mass (i.e., body mass without prosthesis) Normalized VO2 data were averaged to produce 30-s values for minutes 3-6 of each trial. These 30-s values were used in for the described analyses. Both gross and net VO2 were examined, but produced similar results, so only gross VO2 results are reported.

Metabolic Oxygen Consumption (VO2), Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Breath-by-Breath VO2 was collected as participants walked on a treadmill at self selected fast, normal, and slow speeds. Participants walked for 6 minutes at each speed (and in each experimental condition). VO2 was normalized energy expenditure to each participant's biological mass (i.e., body mass without prosthesis) Normalized VO2 data were averaged to produce 30-s values for minutes 3-6 of each trial. These 30-s values were used in for the described analyses. Both gross and net VO2 were examined, but produced similar results, so only gross VO2 results are reported.

Six-minute Walk Test (6MWT), Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Participants were asked to walk as far as possible in 6 minutes. Participants walked around cones placed 30m apart in a level hallway. Total distance (in meters) walked in 6 minutes was recorded.

Walking Speed, Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Walking speed (m/s) was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT).

Borg Rating of Perceived Exertion (CR100), Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Participants were asked to self-report their level of perceived exertion after completing the 6MWT. Participants were shown the Borg CR100 rating scale and asked to identify their level of exertion on a scale ranging from 0 to 120. Higher scores mean greater exertion.

Borg Rating of Perceived Exertion (CR100), Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Participants were asked to self-report their level of perceived exertion after completing the 6MWT. Participants were shown the Borg CR100 rating scale and asked to identify their level of exertion on a scale ranging from 0 to 120. Higher scores mean greater exertion.

Step Length, Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step length (cm) was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT). Step length is reported for both the prosthetic and sound (intact) limbs.

Step Time, Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step time was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT). Step time is reported for both the prosthetic and sound (intact) limbs.

Step Width, Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step width was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT).

Step Time, Participants With Unilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step time was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT). Step time is reported for both the prosthetic and sound (intact) limbs.

Walking Speed, Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Walking speed (m/s) was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT).

Step Width, Participants With Bilateral Amputation

时间窗: Participants were tested after using each foot in their prosthesis for 1 month.

Step width was assessed with a 16-foot CIR GAITRite instrumented walkway while participants performed the 6-minute walk test (6MWT).

次要结局

  • Daily Step Activity, Participants With Bilateral Amputation(Participants' step count data from the 2 weeks prior to in-person testing.)
  • Prosthetic Limb Users Survey of Mobility (PLUS-M), Participants With Unilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Prosthetic Limb Users Survey of Mobility (PLUS-M), Participants With Bilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Daily Step Activity, Participants With Unilateral Amputation(Participants' step count data from the 2 weeks prior to in-person testing.)
  • Patient Reported Outcome Measurement Information System - Fatigue (PROMIS-Fatigue), Participants With Unilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Patient Reported Outcome Measurement Information System - Fatigue (PROMIS-Fatigue), Participants With Bilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Trinity Amputation Prosthesis Experience Scales, Revised (TAPES-R), Participants With Unilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Trinity Amputation Prosthesis Experience Scales, Revised (TAPES-R), Participants With Bilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)
  • Activities-Specific Balance Confidence Scale (ABC), Participants With Unilateral Amputation(Participants were tested after using each foot in their prosthesis for 1 month.)