Bone Remodeling Around HA-coated Acetabular Cups.
试验速览
- 阶段
- 4 期
- 状态
- 已完成
- 发起方
- 入组人数
- 100
- 试验地点
- 1
- 主要终点
- Changes in bone mineral density around the hip prosthesis, measured by DEXA scanning
研究概览
简要总结
This study was designed to investigate the influence of HA-coating on bone remodeling around the cup in cementless THA.
100 patients gave informed consent to participate in a controlled randomized study between porous coated Trilogy® versus Trilogy Calcicoat®. The cup was inserted in press-fit fixation. The femoral component was a cementless porous coated titanium alloy stem (Bi-Metric®), with a modular 28 mm CrCo head. Effect parameters were Harris Hip Score (HHS) and Bone Mineral Density (BMD) determined by DEXA scanning.
Measurements revealed no difference between the two groups after 3 years, neither in clinical outcome nor in terms of periprosthetic bone density. Patients with Body Mass Index above normal regained more bone mineral than patients with normal weight. This finding supports the assumption that load is beneficial to bone remodeling. Advantages of better sealing of the bone-prosthesis interface, preventing polyethylene induced osteolysis, may still be anticipated for the 7 or 12 year follow-up examinations
详细描述
This study was designed to investigate the influence of HA-coating on bone remodeling around the cup in cementless THA.
100 patients gave informed consent to participate in a controlled randomized study between porous coated Trilogy® versus Trilogy Calcicoat®. The cup was inserted in press-fit fixation. The femoral component was a cementless porous coated titanium alloy stem (Bi-Metric®), with a modular 28 mm CrCo head. Effect parameters were Harris Hip Score (HHS) and Bone Mineral Density (BMD) determined by DEXA scanning.
Measurements revealed no difference between the two groups after 3 years, neither in clinical outcome nor in terms of periprosthetic bone density. Patients with Body Mass Index above normal regained more bone mineral than patients with normal weight. This finding supports the assumption that load is beneficial to bone remodeling. Advantages of better sealing of the bone-prosthesis interface, preventing polyethylene induced osteolysis, may still be anticipated for the 7 or 12 year follow-up examinations
研究设计
- 研究类型
- Interventional
- 分配方式
- Randomized
- 干预模型
- Parallel
- 主要目的
- Treatment
- 盲法
- Single (Participant)
入排标准
- 年龄范围
- 18 Years 至 80 Years(Adult, Older Adult)
- 性别
- All
- 接受健康志愿者
- 否
入选标准
- •Patients eligible for cementless THA
排除标准
- •Medical conditions interfering with bone metabolism
结局指标
主要结局
Changes in bone mineral density around the hip prosthesis, measured by DEXA scanning
时间窗: 8-12 years
次要结局
- Changes in hip performance, measured by Harris Hip Score(8-12 years)