Evaluating the Clinical Performance and Usability of Novel Malaria RDTs in Brazil

Not Applicable
Conditions
Malaria
Interventions
Diagnostic Test: NxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and NxTek™ Malaria P.f/P.v RDT
Registration Number
NCT05317117
Lead Sponsor
PATH
Brief Summary

Prospective cross-sectional diagnostic accuracy study with 1,000 patient participants and 30 lay provider/health care worker participants. The patient participant population will be recruited at clinics. At the point of care, study staff will collect capillary blood samples and conduct malaria microscopy (both the standard of care and study specific research-grade microscopy) and two index tests: the NxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and the NxTek™ Malaria P.f/P.v RDT. Venous blood will be collected and transferred to the laboratory where the reference PCR assay and three comparator RDTs will be run. The index RDTs will also be repeated in the laboratory. All clinical management of study participants will follow the standard of care for malaria diagnosis in Brazil and will be based on the standard microscopy result. Confirmatory testing may also be conducted, such as typing and sequencing of Plasmodium genes and antigens of interest, including but not limited to HRP2, HRP3, and pLDH as well as the human inflammatory response marker CRP.

The lay provider/health worker participants will include intended users of the index tests. They will be surveyed to assess index test usability through a questionnaire to assess label and packaging comprehension as well as results interpretation.

Detailed Description

Primary Objectives

1.1 To assess the sensitivity, specificity and, when applicable, positive and negative predicting values (PPV and NPV) \[altogether referred to hereafter as "diagnostic accuracy"\] of NxTek™ Malaria Pf Plus in intended use settings for detecting P. falciparum infections in capillary whole blood samples collected prospectively from patients with symptoms suggestive of malaria.

1.2 To assess the diagnostic accuracy of NxTek™ Malaria Pf/Pv Plus in intended use settings for detecting P. falciparum and P. vivax infections in capillary whole blood samples collected prospectively from patients with symptoms suggestive of malaria.

Secondary Objectives

2.1 To assess the diagnostic accuracy of the study comparator tests in intended use settings for detecting P. falciparum and P. vivax infections in venous whole blood samples collected prospectively from patients with symptoms suggestive of malaria.

1.3 To determine the frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations and assess the impact of those on HRP2-based RDT diagnostic accuracy.

1.4 To assess the comprehension of the test packaging and labeling among intended users (trained lay providers and trained health care workers).

1.5 To assess the ability to read and interpret the test result outputs among intended users (trained lay providers and trained health care workers).

Recruitment & Eligibility

Status
UNKNOWN
Sex
All
Target Recruitment
1030
Inclusion Criteria

Not provided

Exclusion Criteria

Not provided

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
NxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and NxTek™ Malaria P.f/P.v RDTNxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and NxTek™ Malaria P.f/P.v RDTAll participants will be tested with two investigational IVDs at the point of care, the NxTek™ Malaria P.f plus Rapid Diagnostic Test (RDT) and the NxTek™ Malaria P.f/P.v RDT, in addition to comparator tests and the standard of care (microscopy). The investigational tests will not be used to determine any treatment or case management.
Primary Outcome Measures
NameTimeMethod
Diagnostic accuracy of NxTek™ Malaria Pf Plus for the detection of P. falciparum infections5 months

1.1 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of NxTek™ Malaria Pf Plus for the detection of P. falciparum infections in patients with symptoms suggestive of malaria.

Diagnostic accuracy of of NxTek™ Malaria Pf/Pv Plus for the detection of P. falciparum and P.vivax infections5 months

1.2 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of NxTek™ Malaria Pf/Pv Plus for the detection of P. falciparum and P.vivax infections in patients with symptoms suggestive of malaria.

Secondary Outcome Measures
NameTimeMethod
Usability: label comprehension1 week

2.4 Percent of end users who can accurately comprehend key messaging included in the investigational device packaging and labels.

Diagnostic accuracy of the index tests for the detection of P. falciparum infections with HRP2 and/or HRP3 deletions5 months

2.3 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of the index tests for the detection of P. falciparum infections with HRP2 and/or HRP3 deletions in patients with symptoms suggestive of malaria.

Diagnostic accuracy of comparator tests for the detection of P. falciparum and, when relevant, P.vivax infections5 months

2.1 Estimates of diagnostic accuracy characteristics with 95% confidence intervals (sensitivity, specificity, NPV, PPV) of comparator tests for the detection of P. falciparum and, when relevant, P.vivax infections in patients with symptoms suggestive of malaria.

Frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations.5 months

2.2 Frequency of P. falciparum infections containing HRP2 and/or HRP3 mutations.

Usability: results interpretation1 week

2.5 Percent of end users who can accurately interpret the investigational device result output.

Trial Locations

Locations (1)

Centro de Pesquisa em Medicina Tropical de Rondônia (CEPEM)

🇧🇷

Porto Velho, Rondônia, Brazil