A method comparison study of FeNO measurements, between the NOA280i and the NIOX VERO - A method comparison study of FeNO measurements, between the NOA280i and the NIOX VERO

Shizuoka General Hospital0 个研究点目标入组 107 人开始时间: 2016年2月1日最近更新:
适应症

试验速览

阶段
未知
状态
已完成
入组人数
107

研究概览

简要总结

FeNO levels measured using the NIOX VERO (mean +- SD 53.4 +- 38.9 ppb) were significantly higher than those using the NOA 280i (41.7 +- 33.2 ppb; P < .0001). However, there was a significant correlation between FeNO levels measured with the two devices (r = 0.946, P < .0001). The Bland-Altman plot showed a moderate degree of agreement between the 2 devices; the mean inter-device difference was 11.7 ppb (95% limits of agreement -13.5 and 37.0 ppb).

研究设计

研究类型
Interventional

入排标准

年龄范围
20years-old 至 ot applicable(—)
性别
All

入选标准

  • 未提供

排除标准

  • Those who need oxygenation. Those who can not perform technically adequate measurements.