Evaluating Heat Transfer With the Esophageal Cooling Device

Not Applicable
Completed
Conditions
Non-cardiac Surgery
Interventions
Device: Esophageal Cooling
Device: Esophageal Warming
Registration Number
NCT02743884
Lead Sponsor
The Cleveland Clinic
Brief Summary

An esophageal heat exchange tube will be inserted in anesthetized surgical patients. Each patient will have 30 minutes of cooling (circulating fluid at 7 degrees C) and 30 minutes of warming (42 degrees C) with 15-minute recovery break in between. The primary outcome will be heat transfer, determined from inflow and outflow temperatures and fluid flow rate for cooling and warming modes.

Detailed Description

The ECD is made with medical-grade silicone and has 3 ports; 2 coolant ports and a central port. The coolant ports of the ECD are connected to external heat exchanger while a third, central lumen simultaneously allows gastric decompression and drainage. The ECD can be inserted like a standard gastric tube.

The aim of this prospective, interventional study is to quantify heat transfer during warming and cooling using the Esophageal Cooling Device in anesthetized patients having non-cardiac surgery.

An esophageal heat exchange tube will be inserted in anesthetized surgical patients. Each will have 30 minutes of cooling (circulating fluid at 7 degrees C) and 30 minutes of warming (42 degrees C) with 15 -minute break in between; the order of the cooling and warming will be randomized. The primary outcome will be heat transfer, determined from inflow and outflow fluid temperatures and fluid flow rate assessed for both cooling and warming modes.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
19
Inclusion Criteria
  • Elective non-cardiac surgery;
  • Age 18-80 years;
  • Surgery expected to last >2 hours;
  • General anesthesia with endotracheal Intubation;
  • Volatile anesthesia;
  • BMI < 38 kg/m2;
  • ASA physical status 1-3.
Exclusion Criteria
  • Weight < 45 kg;
  • Unstable blood pressure;
  • Known esophageal deformity or evidence of esophageal trauma or esophageal disease that in the opinion of the attending anesthesiologist precludes safe use of ECD.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Esophageal Cooling DeviceEsophageal CoolingEsophageal cooling
Esophageal WarmingEsophageal WarmingEsophageal warming
Primary Outcome Measures
NameTimeMethod
Mean heat transfer at cooling, kcal/h15-60 minutes after intubation during elective non-cardiac surgery

Heat transfer is calculated as difference between inflow and outflow temperatures, multiplied by flow and the specific heat of water

Mean heat transfer at warming, kcal/h15-60 minutes after intubation during elective non-cardiac surgery

Heat transfer is calculated as difference between inflow and outflow temperatures, multiplied by flow and the specific heat of water