临床试验/NCT05186311
NCT05186311已完成不适用

A Prospective Multi-center Study to Evaluate Clinical Equivalence (Method Comparison) Between the BD MiniDraw™ SST™ Capillary Blood Collection Tube and BD MiniDraw™ EDTA Capillary Blood Collection Tube to Respective Comparators

Becton, Dickinson and Company2 个研究点 分布在 1 个国家目标入组 262 人开始时间: 2022年4月22日最近更新:
适应症

试验速览

阶段
不适用
状态
已完成
入组人数
262
试验地点
2
主要终点
Lipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samples

研究概览

简要总结

This study designed to evaluate the safety and effectiveness of a newly developed sample collection device. Participants will be enrolled and samples will be collected by representative future users of the device. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be tested and average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations will be captured and device safety will be assessed.

详细描述

The current method for collection of a large volume of capillary blood (up to 600 µL) involves optional warming of the hand or finger prior to procedure, lancing of the fingertip, squeezing the fingertip by hand, and dripping the blood into an open non-sterile tube. The capillary blood collection process is considered cumbersome and lacks standardization in how much pressure is applied in squeezing of the finger which may lead to poor sample quality or insufficient blood volume. Common sample quality issues for capillary blood collection include hemolysis and dilution of sample with interstitial fluid, potentially causing bias in test results. Open collection also carries a risk of blood exposure for healthcare workers and potential risk of infection at the finger puncture site. A new device, BD Microtainer® Easy Collect Capillary Blood Collection System for the collection of capillary blood was created to standardize the current collection process and improve sample quality.

This study designed to evaluate the safety and effectiveness of the BD Microtainer® Easy Collect Capillary Tube SST™ and BD Microtainer® Easy Collect Capillary Tube EDTA devices. This study will be conducted in two parts, one for the SST product and the other for the EDTA product. A minimum of 100 participants, and a maximum of 150 participants per part will be enrolled at a minimum of three various ancillary healthcare site settings representative of the intended use environment such as retail pharmacy site, patient service center, and/or urgent care clinic. Samples will be collected by representative future users of the BD Microtainer® Easy Collect Capillary Blood Collection System who may not have prior blood collection experience. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be collected, clotted (serum samples only) and centrifuged (serum samples only) before transportation to the laboratory where they will be tested.

Average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations for the SST products will be captured which include hemolysis as measured by Plasma Free Hemoglobin (PFH); collection volume; and duration of sample collection. Visual Observations for the EDTA products will be captured which include frequency of occurrence of instrument flags; frequency of occurrence of platelet clump instrument flags; frequency of instrument platelet clump flags with no platelet clump; evaluate whole blood collection volume; and duration of sample collection. Lastly, device safety will be assessed by device and procedure related adverse events.

研究设计

研究类型
Interventional
分配方式
Na
干预模型
Single Group
主要目的
Other
盲法
None

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者

入选标准

  • Participants ≥18 years of age.
  • Not currently pregnant (self-reported)
  • Adequate access to four fingers (ring and middle finger on each hand) and inner elbows for blood collection procedures
  • Willing and able to comply with all study procedures and evaluations
  • Ability to read, write, and understand English language
  • Provision of signed and dated informed consent form

排除标准

  • Evidence of skin issues such as infections, ulcerations, blisters, peripheral vascular disease, inflammation, extensive scarring or calluses, or healed burns at the fingerstick site. Subject to post-study exclusion period for this or other related studies:
  • A study participant may enroll once in study part A and once in study part B of this clinical study. They may not enroll more than once in a single study part.
  • If a study participant has previously enrolled in one study part and would like to enroll in the second study part, they must wait 8 days from the time of the first study part collection procedures to enroll in the second study part collection procedures.
  • Any condition which, in the opinion of the Investigator, would preclude participation in this study.

结局指标

主要结局

Lipid Analytes and Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mg/dL

Vitamin D (Vit D), in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of ng/mL

Alkaline Phosphatase (ALP), Alanine Transaminase (ALT), and Aspartate Aminotransferase (AST), in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of U/L

White Blood Cells (WBC) and Platelets (PLT) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^3/µL

Mean Corpuscular Volume (MCV) and Red Blood Cell Distribution Width (RDW) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of fL

Chemistry Analytes [Albumin (ALB) and Total Protein (TP)] in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of g/dL

Mean Corpuscular Hemoglobin (MCH) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of pg

White Blood Cell Differential Counts, Hemoglobin A1c, Hematocrit, and Red Blood Cell Distribution Width in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators venous comparator and capillary comparator. This comparison was focused on the analytes with a unit of percentage

Thyroid Stimulating Hormone (TSH) in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of µIU/mL

Metabolic Analytes, in Serum Measured Using the Silica Clot Activator for Serum Samples

时间窗: within 4 hours of sample collection

To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mmol/L

Red Blood Cells (RBC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10\^6/µL

Hemoglobin (HgB) and Mean Corpuscular Hemoglobin Concentration (MCHC) in Plasma Measured With Ethylenediaminetetraacetic Acid (EDTA)

时间窗: within 2 hours of sample collection

To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of g/dL

次要结局

未报告次要终点

研究者

申办方类型
Industry
责任方
Sponsor