MedPath

LarynGuide™ Assisted Tracheal Intubation

Completed
Conditions
Intubation
Interventions
Device: LarynGuide
Registration Number
NCT06397261
Lead Sponsor
Insel Gruppe AG, University Hospital Bern
Brief Summary

This study aims to asses if the application of the larynGuide™, an assistive software running on a video-laryngoscope platform guiding laryngoscopy and intubation advise correctly and reliably on the position of the tracheal tube after an intubation attempt.

Detailed Description

Eligible patients will be prepared for intubation according to the local SOPs of the anesthesia departments with mandatory monitoring of SpO2, HR, NIBP, EtCO2.

For the anesthesia Patients will be pre-oxygenated before induction via face-mask with FiO2 = 1.0 and flow rates of 18L/min until etO2 reaches 90%. Induction of anesthesia and TI will be performed using a combination of sedative/hypnotic drugs, opioids and non-depolarizing muscle relaxant. The following medications will be mandatory as per protocol:

* Hypnotic agent (Propofol 1-4 mg/kg).

* Opioids (Fentanyl 1-3 mcg/kg, Remifentanyl 1-3 mcg/kg).

* Non-depolarizing muscle relaxant (Rocuronium 0.5-1 mg/kg). After induction of anesthesia and the administration of the muscle relaxant, bag-mask ventilation with FiO2 = 1.0 (flow rates of 18 Lmin-1) will be performed until apnea sets in. After induction all patients will be paralyzed to facilitate airway management. Neuromuscular blockade will be assessed by train-of-four monitoring. Thereafter oxygen administration, laryngoscopy and tracheal intubation are performed as usually.

Tracheal intubation performed with the aid of larynGuide™, an assistive software running on a video-laryngoscope platform guiding laryngoscopy and intubation (aiEndoscopic, Zürich, Switzerland). The larynGuide™ will be combined with the C-MAC video laryngoscope (Karl Storz, Tuttlingen, Germany) with a Macintosh nr 3 or 4 blade by a board-certified anesthesiologist. The board-certified anesthesiologist performing the intubation communicates to the research staff when he completed the intubation attempt. The anaesthesiologist in charge (board certified) checks to assess the correctness of the tube position by visual check and by attaching the patient to the ventilator and measuring the endtidal CO2 (Gold standard for detection of tube placement in anesthesia).

For the study the research staff will then - after the confirmation of the anaesthesiologist in charge of the tubes correct placement - look if the software displays a green light or the words "Bad intubation" on the monitor and document either outcome.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
110
Inclusion Criteria
  • male or female patients ≥18 years
  • ASA class I to III
  • undergoing elective surgery in general anesthesia with tracheal intubation
Exclusion Criteria
  • Contraindication for intubation with C-MAC video laryngoscope (Karl Storz, Tuttlingen, Germany) with a Macintosh nr 3 or 4 blade
  • Patients < 18 years
  • No proper trained personnel for the device at the study site
  • Expected impossible mask ventilation.
  • High risk of aspiration (requiring rapid sequence induction intubation)
  • Intracranial surgery
  • Limited knowledge of German language or refusing informed consent

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Arm && Interventions
GroupInterventionDescription
adult patients undergoing elective surgeryLarynGuideadult patients undergoing elective surgery requiring tracheal intubation
Primary Outcome Measures
NameTimeMethod
Rate of correct assessment given by larynGuide™5 minutes

Correctness of the advice measured by accuracy of the LarynGuide™, a software for artificial intelligence assisted tracheal intubation (aiEndoscopic, Zürich, Switzerland) on tube position given at the end of an intubation attempt

Secondary Outcome Measures
NameTimeMethod
first-attempt success rate10 minutes

First-attempt oro-tracheal intubation success rate confirmed with waveform capnography in percentage

Number of incidents of technical problems of larynGuide™10 minutes

Number of the occurrence of any technical problem of the larynGuide™, a software for artificial intelligence assisted tracheal intubation (aiEndoscopic, Zürich, Switzerland)

Number of attempts5 minutes

number of attempts until correct intubation

Rate of Cormack & Lehane Score from 1 to 42 minutes

Cormack-Lehane classification is used to describe laryngeal view during direct laryngoscopy and can be divided into 4 grades of which 1 is the best (all of the glottis can be seen - so the best view) and 4 is the worste (not even the epiglottis is visible - worst view) the occurrence of which we will be counting.

Overall success rate5 minutes

Overall success rate under assistance with larynGuide™

Percentage of glottic opening2 minutes

Best percentage of glottic opening (POGO)

Occurrence rate of airway management complications5 minutes

Complications during and after airway management such oral/airway injury with bleeding, difficult bag-mask ventilation, emesis, aspiration of gastric contents, hypotension necessitating treatment, hypoxia causing bradycardia requiring chest compressions, pneumothorax and the administration of resuscitation drugs

Trial Locations

Locations (1)

Inselspital

🇨🇭

Bern, Switzerland

© Copyright 2025. All Rights Reserved by MedPath