Ankle Syndesmosis Fixation by Antibiotic Releasing Bioabsorbable Screw
- Conditions
- Ankle Fracture
- Interventions
- Procedure: A ciprofloxacin containing bioabsorbable PLGA bone screw
- Registration Number
- NCT01729195
- Lead Sponsor
- Turku University Hospital
- Brief Summary
This single-center randomized trial evaluates the efficacy and safety of a new bone screw (antibiotic releasing bioabsorbable screw) in fixation of syndesmosis in patients with Weber C-type ankle fractures. The comparison is made to subjects treated by conventional metal screw fixation of the syndesmosis. The primary objective is to show that the antibiotic releasing bioabsorbable screw is at least as good as the routinely used metal screw in prevention of syndesmosis widening in patients with Weber C-type ankle fractures. The secondary objective is to show that the clinical outcome of the ankle fracture treatment is equal between patients treated by the antibiotic releasing bioabsorbable screw and the conventional metal screw.
- Detailed Description
This is a randomized parallel-group single centre study. All the patients enrolled will have an acute, closed Weber C-type ankle fracture. The ankle syndesmosis will be fixed using the new medical device (antibiotic containing bioabsorbable screw) or the routinely used device (metal screw). The investigational antibiotic containing bioabsorbable screw for ankle syndesmosis repair (the thread diameter 4.5 mm) will be made of bioabsorbable poly(lactide-co-glycolide) (PLGA) 80:20 and ciprofloxacin, which is a bactericidal antibiotic. The main purpose of the screw is to prevent syndesmosis widening after Weber C-type of ankle fracture. The secondary function of the screw is to reduce the risk for colonization of the implant with bacteria and subsequently prevent biomaterial-related infection. In the control group, a stainless steel screw will be used. All patients will receive the standard of care for concomitant fractures and other injuries of the ankle.
The key variables in the assessment will be radiostereometric (RSA) measurements of the syndesmosis width, CT imaging of the ankle mortise and standard radiographic evaluation of the syndesmosis width and ankle mortise. Clinical outcome of the ankle fracture treatment will be measured using standardized outcome questionnaires (RAND-36, AAOS Foot and Ankle Outcome Instrument, Olerud and Molander score, VAS pain scale). The follow-up of the patients will be 52 weeks. The metal screw will be removed 8 weeks after fracture surgery. The expected mechanical integrity of the bioabsorbable screw is 6 - 8 weeks, while the complete absorption time of the bioabsorbable screw in human body is expected to be approximately two years
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 17
- acute, closed Weber C-type ankle fracture
- adult patients (age 18-70 years)
- female subjects of child-bearing potential must have a negative pregnancy test and an approved contraception for the study duration
- a previous fracture or infection of the injured ankle or the ipsilateral foot
- significant associated soft-tissue injury
- other long-bone fracture of the lower extremities
- documented active infection at any anatomic site
- a known metabolic skeletal disease (osteoporosis or osteomalasia) or a medication affecting bone structure (such as corticosteroid treatment)
- a pathological fracture
- any underlying systemic disease (such as unbalanced diabetes mellitus or rheumatoid disease) or a condition (such as alcoholism or drug abuse) which are known to affect resistance to infection
- history of prosthetic knee or hip replacement
- pregnant women or nursing mothers
- hypersensitivity to fluoroquinolones or related antibiotics (nalidixine acid)
- any other condition that in the judgment of the investigator would prohibit the subject from participating in the study or may hinder the collection of data and interpretation of the results
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description Single-arm A ciprofloxacin containing bioabsorbable PLGA bone screw The syndesmosis injury of the patients will be fixed with a ciprofloxacin containing bioabsorbable PLGA bone screw or a stainless steel metal screw
- Primary Outcome Measures
Name Time Method Radiostereometric analysis of the width of the syndesmosis up to 12 weeks Tantalum markers implanted during surgery into the fibula and tibia. Radiostereometric imaging allows to measure the three-dimensional displacement of the fibula in relation of the tibia.
- Secondary Outcome Measures
Name Time Method Computed tomography in evaluation of the restoration of the normal anatomy Before surgery and repated at 0 and 12 weeks CT imaging allows to evaluate the success in fracture reduction and the maintenance of the restored anatomy during the follow-up
Trial Locations
- Locations (1)
Turku University Hospital
🇫🇮Turku, Finland