Noninvasive Hemoglobin Spot Check Repeatability and Reproducibility in a Blood Donation Setting
Trial Snapshot
- Phase
- Not Applicable
- Status
- Completed
- Sponsor
- Masimo Corporation
- Enrollment
- 20
- Locations
- 1
- Primary Endpoint
- Variation Ratio of Total Repeatability and Reproducibility Hemoglobin Measurements Using Rad-67.
Study Overview
Brief Summary
The primary objective of this clinical investigation is to evaluate the repeatability and reproducibility of using Rad-67 Pulse CO-Oximeter
Study Design
- Study Type
- Interventional
- Allocation
- Na
- Intervention Model
- Single Group
- Primary Purpose
- Other
- Masking
- None
Eligibility Criteria
- Ages
- 18 Years to — (Adult, Older Adult)
- Sex
- All
- Accepts Healthy Volunteers
- Yes
Inclusion Criteria
- •At least 18 years of age
- •Weight at least 110 pounds
- •Subjects with the intention of being screened for eligibility to donate blood
- •The subject has given written informed consent to participate in the study
Exclusion Criteria
- •Subjects with skin abnormalities at the planned application sites that may interfere with sensor application, per directions-for-use (DFU) or trans-illumination of the site, such as burns, scar tissue, infections, abnormalities, etc.
- •Subjects unwilling to remove nail polish or acrylic nails
- •Subjects with blood cancers such as leukemia
- •Subjects with hemoglobin disorders such as sickle-cell anemia and thalassemia or with known history of infectious diseases such as HIV/AIDS, syphilis, hepatitis, etc.
- •Subjects with self-disclosed/known pregnancy at the time of enrollment
- •Subjects deemed not suitable for the study at the discretion of the investigator
- •Subjects unlikely to be able to refrain from excessive motion during data collection. Excessive motion includes postural changes, making hand gestures, involuntary muscular movements, etc.
Arms & Interventions
Test subjects
All subjects are enrolled and receive Rad-67 Pulse Oximeter and DCI Mini sensor for measurement of hemoglobin.
Intervention: Rad-67 Pulse CO-Oximeter and DCI Mini sensor (Device)
Outcomes
Primary Outcomes
Variation Ratio of Total Repeatability and Reproducibility Hemoglobin Measurements Using Rad-67.
Time Frame: 1 hour
Quality of hemoglobin measurements is defined as a variance ratio of total Gage repeatability and reproducibility variance to total variance. Subject, operator and device were considered as components to calculate this ratio.
Secondary Outcomes
No secondary outcomes reported