Spot-Check Noninvasive Pulse CO-Oximeter Validation

Not Applicable
Completed
Conditions
Anemia
Interventions
Device: Masimo Rad-67 Pulse Oximeter
Registration Number
NCT02782806
Lead Sponsor
Masimo Corporation
Brief Summary

In this study, the accuracy of a noninvasive hemoglobin sensor(s) will be assessed by comparison to hemoglobin measurements from a laboratory analyzer.

Detailed Description

Not available

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
368
Inclusion Criteria
  • Male or female
  • Age: Greater than one month
  • Weight: Greater than or equal to 3kg
  • Any racial or ethnic group
  • The subject or the subject's legally authorized representative has given written informed consent/assent to participate in the study
  • If an ICU patient, total hemoglobin of < 9 g/dL at time of screening.
Exclusion Criteria
  • Subjects with skin abnormalities at the planned application sites that may interfere with sensor application, per directions-for-use (DFU) or trans-illumination of the site, such as burns, scar tissue, nail polish, acrylic nails, infections, abnormalities, etc.
  • Subjects deemed not suitable for the study at the discretion of the investigator
  • Subjects unlikely to be able to refrain from excessive motion during data collection. Excessive motion includes postural changes, making hand gestures, involuntary muscular movements, etc.

Study & Design

Study Type
INTERVENTIONAL
Study Design
SINGLE_GROUP
Arm && Interventions
GroupInterventionDescription
Test groupMasimo Rad-67 Pulse OximeterAll subjects are enrolled into the test group and receive Masimo Rad-67 Pulse Oximeter for measurement of hemoglobin.
Primary Outcome Measures
NameTimeMethod
Accuracy of sensorUp to 1 hr per subject

Accuracy willl be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.

Secondary Outcome Measures
NameTimeMethod

Trial Locations

Locations (1)

Masimo Corporation

🇺🇸

Irvine, California, United States