Accuracy of LASIK Flaps Created With the LenSx® Femtosecond Laser - A Pilot Study
- Conditions
- MyopiaHyperopia
- Interventions
- Device: Femtosecond Laser SystemProcedure: LASIK surgery
- Registration Number
- NCT02575911
- Lead Sponsor
- Alcon Research
- Brief Summary
The purpose of this study is to evaluate the thickness accuracy of Laser-assisted in situ keratomileusis (LASIK) flaps created with the LenSx® Laser.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 34
- Diagnosed with myopia or hyperopia and eligible for femtosecond laser initiated LASIK surgery;
- Best corrected visual acuity (BCVA) correctable to at least 20/25 in each eye;
- If contact lens wearer, willing to discontinue contact lens wear prior to screening procedures;
- Willing and able to return for ALL scheduled follow-up examinations;
- Willing and able to provide written informed consent;
- Eligible for bilateral LASIK procedures to be done on the same day;
- Other protocol-specified inclusion criteria may apply.
- Previous corneal surgery, corneal lesions that will impede laser treatment, corneal edema, hypotony, glaucoma, existing corneal implant, or keratoconus;
- Known sensitivity to planned study concomitant medications;
- Participation in any other ophthalmic drug or device clinical study for 30 days prior to or during the time of participation in this study;
- Irregular astigmatism, based on the Investigator's judgement;
- Pregnant, lactating or planning to become pregnant during the course of the study;
- Other protocol-specified exclusion criteria may apply.
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description LenSx LASIK surgery LASIK surgery in both eyes using LenSx® Femtosecond Laser System LenSx Femtosecond Laser System LASIK surgery in both eyes using LenSx® Femtosecond Laser System
- Primary Outcome Measures
Name Time Method Flap Thickness Accuracy Within the Central Zone at Month 3 Postoperative Month 3 postoperative Flap thickness was assessed by optical coherence tomography (OCT) and averaged from 3 separate scans. Flap thickness accuracy was defined as the mean difference between the achieved and desired (desired subtracted from achieved) flap thickness. A positive number represents a postoperative flap thickness that is thicker than the expected flap thickness and vice versa for a negative number. Each eye contributed separately to the analysis.
- Secondary Outcome Measures
Name Time Method Flap Thickness Precision Within the Central Zone at Month 1 and Month 3 Postoperative Month 1, Month 3 postoperative Flap thickness was assessed by OCT and averaged from 3 separate scans. Flap thickness precision (variability of the achieved flap thickness) was defined as the standard deviation of the flap thickness measurement. Each eye contributed separately to the analysis.
Ease of Flap Dissection at Day 0, Operative Day Day 0, operative day Ease of flap dissection (ease of lifting the flap during surgery) was assessed on a scale from 0 to 5 where 0 = Unable to lift flap and 5 = Able to lift flap without any resistance using blunt instrument. Each eye contributed separately to the analysis.
Stromal Bed Quality at Day 0, Operative Day Day 0, operative day Stromal bed quality was assessed by the investigator on a roughness scale from 0 to 5 where 0 = Very rough surface and 5 = Very smooth surface. Each eye contributed separately to the analysis.
Opaque Bubble Layer (OBL) at Day 0, Operative Day Day 0, operative day Opaque bubble layer was assessed by the investigator during the surgery on a scale from 0 to 5 where 0 = No OBL and 5 = 100% OBL in the stromal bed area. Each eye contributed separately to the analysis.
Best Corrected Distance Visual Acuity (BCDVA) by Visit Baseline, Month 1, Month 3 postoperative BCDVA (measurement with the participant's best spectacle correction) was assessed at a distance of 6 meters or 20 feet using a standard Snellen eye chart, both monocularly and binocularly, and reported categorically as a percentage of eyes analyzed.
Manifest Refraction Spherical Equivalent (MRSE) Baseline, Month 1, Month 3 postoperative The participant was manually refracted to his/her best correction using a phoropter and standard Snellen eye charts. Each eye contributed separately to the analysis.
Flap Thickness Accuracy Within the Central Zone at Month 1 Postoperative Month 1 postoperative Flap thickness was assessed by OCT and averaged from 3 separate scans. Flap thickness accuracy was defined as the mean difference between the achieved and desired (desired subtracted from achieved) flap thickness. A positive number represents a postoperative flap thickness that is thicker than the expected flap thickness and vice versa for a negative number. Each eye contributed separately to the analysis.
Uncorrected Distance Visual Acuity (UCDVA) at Month 1 and Month 3 Postoperative Month 1, Month 3 postoperative UCDVA (measurement of uncorrected (without spectacles or other visual corrective devices) distance visual acuity) was assessed at a distance of 6 meters or 20 feet using a standard Snellen eye chart, both monocularly and binocularly, and reported categorically as a percentage of eyes analyzed.
Prediction Error Between Target Versus Achieved Refraction at One and Three Months Post-op Month 1, Month 3 postoperative Prediction error was summarized as a percentage of eyes within 0.5 D and 1.0 D of target at one and three months postoperative