Evaluation of Electrophysiological Parameters related to His Bundle Pacing in Patients With Bradyarrhythmias

Not Applicable
Conditions
Bradyarrhythmias
Registration Number
JPRN-UMIN000031364
Lead Sponsor
Kyorin University School of Medicine Division of Advanced Arrhythmia Management
Brief Summary

Overall, local ventricular thresholds were elevated during follow-up. The distal and proximal HBP thresholds did not differ. The local ventricular thresholds of distal HBP were significantly lower than those of proximal HBP. Optimal values of the ventricular measurements at implantation were determined for predicting the accepted ventricular threshold at 6 months. Distal HBP supported by stable local ventricular capture may be useful for treating pacing-dependent patients.

Detailed Description

Not available

Recruitment & Eligibility

Status
Complete: follow-up complete
Sex
All
Target Recruitment
50
Inclusion Criteria

Not provided

Exclusion Criteria

1) age of equal and more than 90 years-old and of less than 20 years-old; 2) left ventricular ejection fraction of less than 50%.

Study & Design

Study Type
Observational
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
Intracardiac potential and electrohysiological parameters recorded by His bundle pacing lead pre and post placement.
Secondary Outcome Measures
NameTimeMethod
During 12 months after implant 1) Threshold and impedance of His bundle pacing 2) Amplitude of intracardiac potential 3) Complications related to His bundle pacing