Accuracy of a Blood Pressure Measuring Mobile Device in Volunteers
- Conditions
- Blood Pressure
- Interventions
- Device: blood pressure check
- Registration Number
- NCT03382691
- Lead Sponsor
- Seoul National University Hospital
- Brief Summary
The accuracy and safety of blood pressure module of new mobile device (M- HEALTHCARE device) is evaluated according to Korean FDA guidelines in 100 healthy volunteers.
- Detailed Description
Not available
Recruitment & Eligibility
- Status
- COMPLETED
- Sex
- All
- Target Recruitment
- 100
- Participants whose systolic blood pressure <=110 should be at least 10% of whole cohort
- Participants whose systolic blood pressure >=160 should be at least 10% of whole cohort
- Participants whose diastolic blood pressure >=100 should be at least 10% of whole cohort
- Participants whose diastolic blood pressure <=70 should be at least 10% of whole cohort
- Participants whose arm circumference are between 20cm and 30cm should be 25 - 50% of whole cohort
- Participants whose arm circumference are between 30cm and 40cm should be 50 - 75% of whole cohort
- heart disease with abnormal rhythm
- peripheral vascular anomaly or disease
- inadequate to be applied blood pressure cuff (e.g. trauma on arm)
- pregnant or lactating women
Study & Design
- Study Type
- INTERVENTIONAL
- Study Design
- SINGLE_GROUP
- Arm && Interventions
Group Intervention Description EXPERIMENTAL blood pressure check Blood pressure check using mobile device and control device
- Primary Outcome Measures
Name Time Method mean difference of systolic blood pressure 1 day at visit 2 mean difference of blood pressure between new and control device
mean difference of diastolic blood pressure 1 day at visit 2 mean difference of diastolic blood pressure between new and control device
standard deviation of the difference of systolic blood pressure 1 day at visit 2 standard deviation of the difference of systolic blood pressure between new and control device
standard deviation of the difference of diastolic blood pressure 1 day at visit 2 standard deviation of the difference of diastolic blood pressure between new and control device
- Secondary Outcome Measures
Name Time Method
Trial Locations
- Locations (1)
Seoul National University Bundang Hospital
🇰🇷Seongnam Si, Gyenggi DO, Korea, Republic of