FREEDOM - A Frequent Optimization Study Using the QuickOpt Method

Abbott Medical Devices2 个研究点 分布在 1 个国家目标入组 1,647 人开始时间: 2006年10月1日最近更新:
适应症
干预措施

试验速览

阶段
不适用
状态
已完成
入组人数
1,647
试验地点
2
主要终点
Heart Failure Clinical Composite Score

研究概览

简要总结

The objective of this study is to demonstrate that frequent atrio-ventricular (AV/PV) and inter-ventricular (V-V) delay optimization using QuickOpt in patients with cardiac resynchronization therapy device results in improved clinical response over standard of care (i.e. empiric programming or one-time optimization using any non-intracardiac electrogram optimization methods).

详细描述

  • This is a prospective, double-blinded, multicenter, randomized study
  • Patient could be enrolled up to 2 weeks post CRT-D implant and are followed for 12 months post implant with follow-up visits at 3, 6, 9 and 12 months
  • Patients will be randomized at enrollment to either Group 1 ("QuickOpt Group") or Group 2 ("Control Group").
  • Group 1 - The patient's device is programmed to sequential biventricular pacing mode with AV/PV and VV delays optimized using QuickOpt. For Group 1 patients, optimization using QuickOpt is performed at enrollment, 3 month, 6 month, 9 month, 12 month and at any unscheduled follow-up visits.
  • Group 2 - The patient's device is programmed to either simultaneous or sequential BiV pacing mode as per physician's discretion. The AV/PV and inter-ventricular (VV) delays could be programmed empirically or optimized using any non-IEGM based method as per sites standard of care. However, the Group 2 patients can be optimized only once within the first 4 weeks post implant. Any AV/PV and VV delay optimizations performed after 4 weeks post implant in Group 2 patients will be considered protocol deviations.

研究设计

研究类型
Interventional
分配方式
Randomized
干预模型
Parallel
主要目的
Treatment
盲法
Double (Participant, Outcomes Assessor)

入排标准

年龄范围
18 Years 至 —(Adult, Older Adult)
性别
All
接受健康志愿者
否

入选标准

  • •Patient meets current CRT-D indications and be implanted with a St. Jude Medical (SJM) CRT¬D device with VV timing and a compatible lead system.
  • •Patient has the ability to complete a 6-minute hall walk with the only limiting factor to be fatigue or shortness of breath.
  • •Patient has the ability to independently comprehend and complete a QOL questionnaire.

排除标准

  • •Patient has an epicardial ventricular lead system.
  • •Patient has the ability to walk ≥ 450 meters in 6 minutes
  • •Patient has limited intrinsic atrial activity (≤ 40 bpm).
  • •Patient has persistent or permanent atrial fibrillation (AF).
  • •Patient has a 2° or 3° heart block.
  • •Patient's life expectancy is less than 1 year.
  • •Patient is pregnant.
  • •Patient is on IV inotropic agents.

研究组 & 干预措施

QuickOpt (Treatment)

Experimental

Frequent optimization using QuickOpt to optimize the AV/PV and VV Delays.

干预措施: QuickOpt (Device)

Control

Active Comparator

Empiric programming or one-time optimization using a non-IEGM method.

干预措施: Control (Device)

结局指标

主要结局

Heart Failure Clinical Composite Score

时间窗: 12 months

The clinical composite score classifies each randomized patient as improved, unchanged, or worse depending on the clinical response during and the clinical status at the end of the trial. Patients are considered improved if at the final visit they experienced a favorable change in NYHA functional class or in the patient global assessment (or both) but did not experience any major adverse clinical events during the course of the trial. Patients are considered worse if they experienced a major clinical event during the study duration or reported worsening of their NYHA class or global assessment at the final visit. Patients are considered unchanged if they are neither improved nor worse.

次要结局

  • All Cause, Cardiovascular and Heart Failure Hospitalization(12 months)
  • All-cause, Cardiovascular and Heart Failure Mortality;(12 months)