MedPath

BDx 7 Day ECG Patch: Electrode Adhesion Clinical Study

Completed
Conditions
ECG Monitoring
Cardiac Patients
Registration Number
NCT02133261
Lead Sponsor
Bardy Diagnostics, Inc.
Brief Summary

The purpose of this research study is to learn if a new 7 day patch ECG recording monitor can serve the same purpose as standard monitoring systems. The new 7day patch is significantly smaller and lighter than previous ECG recorders.

Detailed Description

The primary purpose of this research is to evaluate ECG signal quality of the BDx 7 day ECG Patch heart monitor after 7 full days of wear, establishing that the device is safe and effective to wear for up to 7 days.

Recruitment & Eligibility

Status
COMPLETED
Sex
All
Target Recruitment
10
Inclusion Criteria
  • cardiac patient volunteers (5 male and 5 female)
Exclusion Criteria
  • skin rash or infection over the sternum
  • Sternal incision within 3 months from the date of enrollment

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
adhesive performance/durability7 days

The study will evaluate adhesive performance/durability by evaluating the Signal Quality of ECG data on the subjects (5 men and 5 women) in accordance with the requirements of FDA's ECG Electrode Guidance (section 5.A.3)1 and the ANSI/AAMI EC-12 Disposable ECG electrode technical standard (sections 4.4 and 5.4)2.

Secondary Outcome Measures
NameTimeMethod
human factors7 days

Skin comfort or discomfort Device comfort Device stability and contact

Trial Locations

Locations (1)

EvergreenHealth

🇺🇸

Enumclaw, Washington, United States