Skip to main content
Clinical Trials/NCT04562584
NCT04562584CompletedNot Applicable

Accuracy of Pulse Oximeters With Profound Hypoxia

MyHomeDoc Ltd.2 sites in 1 country12 target enrollmentStarted: June 3, 2020Last updated:
Conditions

Trial Snapshot

Phase
Not Applicable
Status
Completed
Sponsor
Enrollment
12
Locations
2
Primary Endpoint
Saturation level measurement

Study Overview

Brief Summary

The aim of this project is to test the accuracy of pulse oximeters (devices intended to measure oxygen saturation in blood noninvasively) during mild, moderate and severe hypoxia (a lower-than-normal concentration of oxygen in arterial blood); ie, a range of arterial HbO2 saturations from 100 to down to 70%. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin (hemoglobin carrying oxygen) saturation, that is arterial blood sample processed in a laboratory.

Study Design

Study Type
Interventional
Allocation
Na
Intervention Model
Single Group
Primary Purpose
Diagnostic
Masking
None

Eligibility Criteria

Ages
18 Years to 50 Years (Adult)
Sex
All
Accepts Healthy Volunteers
Yes

Inclusion Criteria

  • The subject is male or female, aged ≥18 and <
  • The subject is in good general health with no evidence of any medical problems.
  • The subject is fluent in both written and spoken English.
  • The subject has provided informed consent and is willing to comply with the study procedures.

Exclusion Criteria

  • The subject is obese (BMI>30).
  • The subject has a known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP.
  • Subject has diabetes.
  • Subject has a clotting disorder.
  • The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
  • The subject has any other serious systemic illness.
  • The subject is a current smoker.
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
  • The subject has a history of fainting or vasovagal response.
  • The subject has a history of sensitivity to local anesthesia.
  • The subject has a diagnosis of Raynaud's disease.
  • The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • The subject is pregnant, lactating or trying to get pregnant.
  • The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
  • The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study.

Outcomes

Primary Outcomes

Saturation level measurement

Time Frame: 1 hour

Induction of hypoxia by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide to allow measurement of oxyhemoglobin saturation between 70-100%. Each plateau level of oxyhemoglobin saturation was maintained for at least 30 seconds or until reference pulse oximeters readings were stable

Secondary Outcomes

No secondary outcomes reported

Investigators

Sponsor
MyHomeDoc Ltd.
Sponsor Class
Industry
Responsible Party
Sponsor

Study Sites (2)

Loading locations...

Similar Trials