Comparative STudy of the ARGos Biometer Compared to UltraSound Biometer in B-mode in Case of Measure Failure With a the Conventional Biometer

Not yet recruiting
Conditions
Cataract
Registration Number
NCT05534373
Lead Sponsor
Assistance Publique - Hôpitaux de Paris
Brief Summary

Biometry has become one of the most important steps in modern cataract surgery. The axial length of the eye may be measured by ultrasound (either contact or immersion) or by optical means.

Non-contact optical biometry has become the gold-standard because of its ease of use, accuracy, and reproducibility. However, the main disadvantage of the optical methods is their inability to obtain axial length measurements in approximately 10% of eyes, typically those with dense posterior subcapsular cataracts.

By Using SS-OCT device, it has been reported that the AL measurements cannot be made in 0.6 to 7.4% of cases . Therefore, measurement by US biometer is still necessary for advanced cataract. The ultrasound measure the LA between the corneal epithelium and the vitreomacular interface. Ultrasound in B mode (or Brilliance) with the help of the control vector allows a 2-dimensional control of the plan used for measurement. This method is used in current practice in our centers when LA measurements is not possible with our optical SS-biometer.

Recently Alcon has integrated the ARGOS® biometer's diagnostic testing with Alcon's Cataract Refractive Suite technologies. ARGOS® is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS® measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White-to-White (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool. ARGOS® includes an operation mode for when the retina is hardly or not visible at all denominated the "Enhanced Retinal Visualization mode" (ERV) that allows, in most cases, visualization of the retina. This signal in the retinal region is enhanced by shifting the coherence function (Argos Users manual).

In a prospective observational study conducted in 2 centers in USA involving 107 eyes, they have evaluated the repeatability and reproducibility of ARGOS® measurements in comparison with IOLMaster®500 and LENSTAR LS 900. They have shown that the AL acquisition rate was higher with ARGOS® (96%) compared with both LENSTAR LS 900 (79%) and IOLMaster®500 (77%). The reason was that ARGOS® was more successful at measuring AL in eyes with dense cataract.

Detailed Description

Our recent experience with Argos has shown that LA measurements were even possible with our without ERV mode in case of very dense (ie: cataracts in diabetic patients) or posterior subcapsular cataracts.

These observations lead us to propose a retrospective study to compare Argos to our gold standard "ultrasound biometer in mode B" for LA measurements in case of failure with the conventional optical biometer used in our current practice (Lenstar LS900, IOLMaster 500, 700) In current practice, the biometry measurements are conducted with our conventional optical biometer during the visit prior to cataract surgery.

At the same time, measures will be conducted on Argos with or without the ERV mode If axial length measurements are impossible with our conventional optical biometer due to the level of cataract density, axial length measurements will be conducted with our ultrasound biometer in mode-B (Quantel Medical, Aviso S with the 10MHz probe) according to our current practice.

We will retrospectively collect the consecutive data from eyes for which axial length measurements with our conventional biometer were not possible

Recruitment & Eligibility

Status
NOT_YET_RECRUITING
Sex
All
Target Recruitment
110
Inclusion Criteria
  • Patient over 18 years of age
  • All consecutive patient measured on Argos and any other conventional optical biometer and presenting a failure with one of the conventional optical biometer in AL measurement due to cataract density
Exclusion Criteria
  • Silicon tamponade in the posterior segment
  • Patients who are informed and objected to participating to the research
  • Patient subject to legal protection measures.

Study & Design

Study Type
OBSERVATIONAL
Study Design
Not specified
Primary Outcome Measures
NameTimeMethod
To describe axial length (AL) measured by the Argos biometer in case of measure failure with a conventional optical biometer and to compare it with the AL measured with ultrasound biometry in B-mode (considered to be the gold standard)2 months

Axial length (AL) in millimeters obtained with Argos and ultrasound biometer in B-mode scans (considered to be the reference method for dense cataract)

Secondary Outcome Measures
NameTimeMethod
to describe Pachymetry in micrometers obtained with Argos biometer, conventional biometer and ultrasound biometer2 months

Pachymetry in micrometers obtained with Argos biometer, conventional biometer and ultrasound biometer

to measure anterior chamber depth in millimeters obtained with Argos biometer, conventional biometer and ultrasound biometer2 months

Anterior chamber depth in millimeters obtained with Argos biometer, conventional biometer and ultrasound biometer

to measure lens thickness obtained with Argos biometer and ultrasound biometer2 months

Lens thickness obtained with Argos biometer and ultrasound biometer